Enhancing Lives, Delivering Comfort O P E R A ® F L O A U T O M A T T R E S S S Y S T E M Instructions and Technical Specifications V I S I T O P E R A B E D S . C O M C A L L 0 3 3 3 2 2 2 8 5 8 4...
Mattress and Power Unit Installations Program Settings Patient Transfer and Transport Emergency CPR Deflation Cleaning Instruction Routine Maintenance and Storage Troubleshooting Return for Service Warranty Product Specifications 1 | Opera Beds | Opera Flo Auto Mattress System Instructions and Technical Specifications | Version 2...
Page 3
Use of genuine components is essential for optimal performance. If you do not fully understand all the instructions, safety precautions, and warnings, do not use this device. If you have questions, please contact Opera on 0333 222 8584. operabeds.com | 0333 222 8584 | 2...
This product must be disposed of in a separate refuse collection in the European Union. Do not dispose of as normal domestic waste. Read instructions. Manufacturer. Date of Manufacture. 3 | Opera Beds | Opera Flo Auto Mattress System Instructions and Technical Specifications | Version 2...
2. Safety Precautions 2.1 Proper Operation To ensure proper operation, please inspect and verify that all parts are set up properly and are anchored securely to the bed system. Verify that the mattress replacement system does not interfere with the bed frame function. Do not place anything on top of the power unit.
For equipment that is no longer functional, make sure to follow national, state, and/or facility requirements for disposal of the unit. 5 | Opera Beds | Opera Flo Auto Mattress System Instructions and Technical Specifications | Version 2...
3.7 Maximum Operating Temperature The air mattress operates at a maximum temperature of 100 °F (37.7 °C). 3.8 Sharp Protrusions Ensure that there are no protruding objects, sharp points, or bed springs under the mattress as these could puncture the air cells. 3.9 Disposal of Waste This device must not be disposed of as normal household waste after the end of its service life.
Audible alarm will sound for 15 seconds at each cycle and the indicator will stay on until the problem has been solved. 7 | Opera Beds | Opera Flo Auto Mattress System Instructions and Technical Specifications | Version 2...
Page 9
d. Comfort Adjustment Simply press the button to adjust the patient’s comfort from -2 to 2. Please adjust the comfort setting if the mattress is too soft or firm to suit each patient. Caregivers should always perform a hand check by placing their hands underneath patient’s pelvis area to check if there is sufficient air support to ensure the patient is not bottoming out.
Secure the power unit on to the foot board using hanging hooks. 6.4 Connect Hose Couplers Firmly connect the air hose couplers to the couplers on the power unit. 9 | Opera Beds | Opera Flo Auto Mattress System Instructions and Technical Specifications | Version 2...
6.5 Check CPR Pull Cord Unzip mattress cover and inspect CPR latch is fully inserted into the socket. It should arrive fully inserted when first purchased, but it is worthwhile checking, especially when re-installing the mattress. 6.6 Power Up Unit Plug the power unit into an electrical outlet, and turn on the main power rocker switch on the right side of the power unit.
To resume normal operation, please follow the instructions beginning in Section 6.2. 11 | Opera Beds | Opera Flo Auto Mattress System Instructions and Technical Specifications | Version 2...
9. Emergency CPR Deflation In the case of emergency, pull hard to remove the mattress’ CPR latch and detach the quick release coupling from the control unit for DANGER emergency deflation and turn off the power unit by pressing the power/ mute button on the control panel at the same time for emergency deflation.
30 sec. after system 222 8584 for assistance. starts. Alert LED is lit during Check if there is leakage in air operation. tubes* or air cells. 13 | Opera Beds | Opera Flo Auto Mattress System Instructions and Technical Specifications | Version 2...
Page 15
Check if air cells are cut or Please contact Opera on 0333 cracked 222 8584 for assistance. Alternating or Please contact Opera on 0333 Constant mode is Possible control failure.
15. Product Specifications System Name Flo Auto Mattress System Power Unit (Air Pump) Model SR366 2000 x 900 x 180mm (7” depth) Mattress Dimension 2000 x 1050 x 180mm (7” depth) Number of Air Cells 20 Cells Power Control Unit...
Page 18
Notes 17 | Opera Beds | Opera Flo Auto Mattress System Instructions and Technical Specifications | Version 2...