Operating instructions CellaTemp PX 41
11.11 Damping function
When the target object's temperature is erratic, the damping function smoothens these
temperature fluctuations in order to stabilize the measuring signal� The greater the time
constant, the lower the effect of fluctuations on the yielded temperature reading�
Two-colour mode [qu] → [q. F iL]
One-colour mode [LI] → [' . F iL] or [L2] → [" . F iL]
Ao
[mA]
1
Output signal without smoothing function
2
Output signal with smoothing function
11.12 Peak hold function
It might often be desirable to determine the peak temperature during a defined
time period, for example when the objects to be measured move past the pyrom-
eter, resulting in temperature readings which would appear to be cyclical� In this
mode, the displayed temperature reading will not drop between targeted objects�
The peak temperature reading will be held for a preset time period�
The hold time [t iME] can be set from 0�1 to 999�9 sec� The maximum
temperatures sampled during the defined hold time will be saved� It make sense to
1
2
25
EN
t