Toshiba LF620 Instruction Manual page 88

Electromagnetic flowmeter converter
Hide thumbs Also See for LF620:
Table of Contents

Advertisement

You can check or change the count rate and pulse width as described below.
The following is an example to change the count rate from 0.01m
Switch operation
STEP1 
 
 
 
STEP2 
 
 
 
STEP3 
 
 
 
STEP4 
 
 
 
STEP4(=END) 
 
 
 
To set the pulse width setting mode or pulse width, select the relevant item below from the setting
menu.
Pulse width setting mode
Pulse width
Display example
CNT RATE 
The currently set count rate (0.01m
 
example) appears.
  0.01000 
 
Then push
3
        m
 
 
 
 
ESC 
 
CNT RATE 
The switches at the bottom change.
 
(
  0.01000 
At the same time, the cursor appears.
 
 
3
        m
 
 
 
 
 
CNT RATE 
Push
 
of the unit and push
  10.0000 
"m
 
 
        L 
 
 
 
CNT RATE 
Push
 
you want to change and push
  0.90000 
change the number of the digit.
 
 
        L 
 
 
 
CNT RATE 
When you push
 
to confirm whether the setting is OK or not.
  0.90000 
If OK, push
 
setting, push
        L 
SET OK? 
Pushing
 
and exits the setting screen.
ESC 
O K
N O
 
PLS MODE
PLS WID
− 87 −
6 F 8 A 0 9 2 7  
3
to 0.9 L.
Description
.
are shown.)
to move the cursor to the digit
3
" to "L".
to move the cursor to the digit
, a message appears
. If you want to redo the
O K
.
N O
cancels the setting operation
ESC
3
in this
to change from
to