Quick Start Guide
Install retractable sensor with an insertion adapter
Figure 2-4: Insertion adapter (PN 23166-00 or PN 23166-01) with
Rosemount retractable sensor
A. 1-in. NPT
B. Ferrule fitting process connector assembly: PN 23166-00 or PN
23166-01
C. Sensor body
Procedure
1. Slide the fitting onto the sensor from the back of the sensor.
Install the threads facing forward for insertion installations.
Install the threads facing backward for connection to a
submersion pipe.
2. Connect the NPT threads on the adapter to the process piping
or submersion pipe.
3. Adjust the position of the fitting on the sensor to achieve the
desired sensor insertion length.
4. While holding the sensor in position, tighten the hex nut of the
male connector to firmly secure the sensor in place.
NOTICE
Over-tightening the hex nut may damage the ferrule.
When you tighten the hex nut, the PTFE ferrule inside the
compression fitting clamps the sensor tube.
8