Periodic Inspection Checklist | Ec-Track System, Gantry And Robin - Invacare Robin Service Manual

Ceiling hoist
Hide thumbs Also See for Robin:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

5.3 Periodic Inspection Checklist | EC-Track System, Gantry and Robin

In accordance with ISO 10535.
CAUTION!
Risk of injury and damage to property
Failure to ensure an adequate Periodic Inspection of the product can compromise the safety of people and
property.
– A periodic inspection of the product must be performed at least every 12 months unless otherwise stated
in local requirements.
– The frequency of inspection must be increased if the product is continuously exposed to high humidity, high
condensation and corrosives (e.g., chlorine or ammonia gases), to prevent product impairment.
– Inspections MUST be performed by a qualified person who is acquainted with the design, use and care of
the product.
– The inspection record MUST be recorded in the logbook and reported to the owner.
Is the EC-Track on the installation?
□ Yes (Complete the following section):
The Invacare EC-Track System installation label shows the following information:
Maximum SWL
Label Number (LN)
□ No (Proceed to 2. Checkpoints | Robin).
1. Checkpoints | EC-Track and Gantry
1.1 The installation label is present, visible and readable.
1.2 All connection points, brackets to structure are tightened with
recommended torque.
Retighten, if necessary (if NOT accessible please see point 1.21 on load test).
1.3 All connection points, brackets to rail system are tightened with
recommended torque. If the system is the Gantry, all the connections are
tightened with recommended torque.
Retighten, if necessary (if NOT accessible please see point 1.21 on load test).
1.4 All rail ends are secured with end stops, sufficiently tightened.
Retighten, if necessary.
1.5 End stops don't show any sign of damage, wear and/or deformation.
1.6 Brackets don't show any sign of damage, wear and/or deformation. If
the system is Gantry, connections don't show any sign of damage, wear
and/or deformation.
1.7 Trolleys don't show any sign of damage, wear and/or deformation.
1.8 The trolley (and if applicable, traverse trolleys) run smoothly and silently
in the entire track system.
1.9 Check and adjust if required the trolley friction brake.
1.10 End Covers are assembled (If applicable).
1.11 Rail systems with track connections: The connections are aligned.
Re-align, if needed.
Transit Coupling / Track Switch (if applicable)
1.12 Lock fixtures all OK (secure to tracking, each claw moves independently,
correctly aligned with tracking slots, no obstructions to prevent correct
function, safety plate is fitted).
1.13 Check traverse trolley mounting fixtures.
1586747-F
Installation Date
ü û
Comments
Servicing
11

Advertisement

Table of Contents
loading

This manual is also suitable for:

Robin mover

Table of Contents