Operation - Graco 970159 Instructions-Parts List Manual

55 gallon, 20:1 king module supply units
Table of Contents

Advertisement

Pressure Relief Procedure
WARNING
SKIN INJECTION HAZARD
The system pressure must be manually
relieved to prevent the system from
starting or spraying accidentally. Fluid
under high pressure can be injected through the
skin and cause serious injury. To reduce the risk of
an injury from injection, splashing fluid, or moving
parts, follow the Pressure Relief Procedure
whenever you:
D are instructed to relieve the pressure,
D stop dispensing,
D check or service any of the system equipment,
D or install or clean the dispense ball valve.
1. Relieve system pressure to which Graco module
supply units are attached.
2. Close dispenser ball valves (S). Unhook ball valves
from system. See Fig. 1.
3. Close pump air bleed valve (H, required in your
system). See Fig. 1 or Fig. 7.
4. Shut off main air bleed valve (E, required in your
system). Set ram director valve (U) to DOWN.
Ram will slowly drop.
5. Hold metal part of dispenser ball valve firmly to
side of grounded metal pail, and open ball valve to
relieve pressure.
6. Close dispenser ball valve.
7. Repeat procedure for each ram (for models
970159 and 970160 only).
12
310334

Operation

Flushing the System
Flush the system before initial use to prevent material
contamination.
Flush the system before performing the initial
material loading procedure. The system was fac-
tory-tested using a light soluble oil, a soybean oil, or
other oil as tagged. Flush the system to avoid con-
taminating the material that has been designated for
initial material loading.
To flush the system, perform the following procedure:
1. Select material to be used.
Use fluids and solvents that are chemically compat-
ible with the equipment wetted parts. Refer to the
Technical Data sections of all the equipment
manuals. Always read the material manufacturer's
literature before using fluid or solvent in this pump.
2. Verify that factory-test oil and material are compat-
ible:
a. If substances are compatible, omit following
step b and proceed to step 4.
b. If substances are incompatible, perform re-
maining steps to flush system.
3. Select drum of solvent that dissolves, cleans, and
eliminates factory-test oil from system. If neces-
sary, check with Graco Distributor or material sup-
plier for recommended solvent.
4. Before flushing, be sure entire system and flushing
drums are properly grounded. Refer to System
Grounding, on page 8.
5. Using compatible solvent, perform Starting and
Adjusting System Rams procedure and Starting
and Adjusting the Pump procedure on page 14.
6. Flush solvent through system for 1 to 2 minutes.
7. Remove drum containing solvent.
CAUTION
WARNING

Advertisement

Table of Contents
loading

This manual is also suitable for:

288649970160970163

Table of Contents