Endress+Hauser Cerabar M PMC51 Operating Instructions Manual page 16

Process pressure measurement
Hide thumbs Also See for Cerabar M PMC51:
Table of Contents

Advertisement

Installation
16
3.3.6
Assembling and mounting the "separate housing" version
1
Fig. 12:
"Separate housing" version
1
In the case of the "separate housing" version, the sensor is delivered with the process connection and cable ready
mounted.
2
Cable with connection jack
3
Pressure compensation
4
Connector
5
Locking screw
6
Housing mounted with housing adapter, included
7
Mounting bracket suitable for wall and pipe mounting, included
Assembly and mounting
1.
Insert the connector (item 4) into the corresponding connection jack of the cable (item 2).
2.
Plug the cable into the housing adapter (item 6).
3.
Tighten the locking screw (item 5).
4.
Mount the housing on a wall or pipe using the mounting bracket (item 7).
When mounting on a pipe, tighten the nuts on the bracket uniformly with a torque of at least
5 Nm (3.69 lbs ft).
Mount the cable with a bending radius (r) 120 mm (4.72 in).
6
5
4
3
2
r ³ 120 mm
Cerabar M Analog
7
P01-XMx5xxxx-11-xx-xx-xx-009
Endress+Hauser