Co 2 Parameter Settings - E+E Elektronik EE895 User Manual

Miniature sensor module for co2, temperature and barometric pressure
Hide thumbs Also See for EE895:
Table of Contents

Advertisement

4.5.3.
CO
Parameter Settings
2
The CO
parameter settings are non-volatile since they are stored in the flash memory.
2
Function
address
code
0x06
1) Register address starts from 0
Tab. 18
CO
parameter settings
2
*) The EE895 features an exponential moving average filter which to a certain degree suppresses the influence
of short-term CO
response time of the EE895. A higher filter coefficient leads to smoother output data and to longer response time,
see figure Fig. 12. Fig. 13 shows the number of samples required to reach 63% or 90% of a CO
of the filter coefficient.
Fig. 12
Step response vs. samples
Register
1)
Description
[HEX]
0x1450
CO
measuring interval
2
0x1451
CO
filter coefficient*)
2
0x1452
CO
customer offset
2
variations onto the output CO
2
Unit
Note
16 bit integer unsigned
Scale 1:10. Example: 150 = 15 seconds
S
Range 100...36 000
Default 15 s
16 bit integer unsigned
-
Range 1...20
Default: 4
16 bit integer signed
ppm
Range -32 786...32 785
Default: 0 ppm
data. The filter coefficient is user selectable and affects the
2
Miniature Sensor Module for CO
User Manual EE895
step as function
2
, Temperature and Barometric Pressure | 20
2

Advertisement