E+E Elektronik EE1900 User Manual

E+E Elektronik EE1900 User Manual

Humidity module for climate chambers
Hide thumbs Also See for EE1900:

Advertisement

Quick Links

User Manual
EE1900
HUMIDITY MODULE FOR
CLIMATE CHAMBERS
BA_EE1900 // v1.1 // Modification rights reserved

Advertisement

Table of Contents
loading

Summary of Contents for E+E Elektronik EE1900

  • Page 1 User Manual EE1900 HUMIDITY MODULE FOR CLIMATE CHAMBERS BA_EE1900 // v1.1 // Modification rights reserved...
  • Page 2 E+E Elektronik Ges.m.b.H. does not accept warranty and liability claims neither upon this publication nor in case of improper treatment of the described products. The document may contain technical inaccuracies and typographical errors. The content will be revised on a regular basis.
  • Page 3: Table Of Contents

    CONTENT General ..............................4 Explanation of Symbols ..........................4 Safety Instructions ............................4 1.2.1 General Safety Instructions ..........................4 1.2.2 Mounting, Start-up and Operation ........................4 1.2.3 Intended Use ................................5 Environmental Aspects ..........................5 ESD Protection .............................5 Scope of Supply ............................5 Product Description ..........................5 General ................................5 2.1.1 Sensing Probes ..............................6 2.1.2...
  • Page 4: General

    This symbol indicates safety information. It is essential that all safety information is strictly observed. Failure to comply with this information can lead to personal injuries or damage to property. E+E Elektronik® assumes no liability if this happens. This symbol indicates instructions.
  • Page 5: Intended Use

    The sensing probe and the electronics of the EE1900 undergo the factory adjustment together and may not be separated from each other. Do not cut, shorten or extend the probe cable. The analogue output of EE1900 can be set to voltage or current with a slide switche on the electronics board (see chapter 4.1).
  • Page 6: Sensing Probes

    2.1.1 Sensing Probes The EE1900 features a plastic (PPS) probe according to the order which can be employed over the entire T range according to the technical data. 2.1.2 Filter Caps The right choice of filter cap is very important for optimal performance under certain environment conditions.
  • Page 7: Electrical Connection

    Use only power supply units according to EN 61140, protection class III (EU) and protection class II (North America). For performance according to the specifications, the electronics board must be properly grounded, see chapter 4 (No. 9 “Grounding“ - Fig. 3). User Manual EE1900 Humidity Module for Climate Chambers...
  • Page 8: Setup And Configuration

    EE-PCS Product Configu- ration Software, free download from www.epluse.com/configurator. Use the optional configuration cable HA011017 to connect the EE1900 to the USB port of the PC. User Manual EE1900 Humidity Module for Climate Chambers...
  • Page 9: Error Indication On The Analogue Output (Namur)

    Error Indication on the Analogue Output (NAMUR) The EE1900 features an error indication on the analogue output according to NAMUR recommendations. A device failure is indicated by 11 V in case of voltage output and by 21 mA in case of current output.
  • Page 10 (for instance more than 6 m/s with the stainless steel grid filter), the required sensor temperature might not be reached. In such a case, the heating process is automatically interrupted after 2 minutes. User Manual EE1900 Humidity Module for Climate Chambers...
  • Page 11: Calibration / Adjustment

    An adjustment can be followed by a calibration which documents the accuracy of the adjusted specimen. The adjustment of EE1900 can be performed with the EE-PCS Product Configuration Software, free download from www.epluse.com/configurator.
  • Page 12: Replacement Parts / Accessories

    The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k=2 (2-times standard deviation). The accuracy was calculated in accordance with EA-4/02 and with regard to GUM (Guide to the Expression of Uncertainty in Measurement). 2) Other filters see data sheet “Accessories”. User Manual EE1900 Humidity Module for Climate Chambers...
  • Page 13 HEADQUARTERS SUBSIDIARIES E+E Elektronik Ges.m.b.H. E+E Elektronik China E+E Elektronik Germany E+E Elektronik Korea Langwiesen 7 18F, Kaidi Financial Building, Schöne Aussicht 8 C Suite 2001, Heungdeok IT 4209 Engerwitzdorf No.1088 XiangYin Road 61348 Bad Homburg Valley Towerdong, 13, Austria 200433 Shanghai Tel.:...