6.2.4

Mounting the sensor

WARNING
L
Risk of injury when mounting the sensors and strapping bands!
Due to the increased risk of cuts, wear suitable gloves and protective goggles.
Sensor configuration and settings
DN 15 to 65 (½ to 2½")
Strapping band
2 traverses
[mm (in)]
1)
Sensor distance
Sensor distance
Wire length →  35
1)
Depends on the conditions at the measuring point (measuring pipe, medium etc.). The dimension can be
determined via FieldCare or Applicator. See also the Result sensor distance / measuring aid parameter in
the Measuring point submenu
2)
Only up to DN 600 (24")
Mounting types
Sensor holder with U-shaped screws
Can be used for
• Measuring devices with measuring range DN 15 to 65 (½ to 2½")
• Mounting on pipes DN 15 to 32 (½ to 1¼")
Procedure:
1. Disconnect the sensor from the sensor holder.
2. Position the sensor holder on the measuring pipe.
3. Fit the U-shaped screws through the sensor holder and lightly grease the thread.
4. Screw the nuts onto the U-shaped screws.
5. Position the sensor holder correctly and tighten the nuts uniformly.
 15
Holder with U-shaped screws
1
Sensor holder
DN 50 to 4000 (2 to 160")
Strapping band
1 traverse
2 traverses
[mm (in)]
[mm (in)]
1)
Sensor distance
Measuring rail
1
Installation
Welded bolt
1 traverse
2 traverses
[mm (in)]
[mm (in)]
1)
1)
Sensor distance
Sensor distance
1) 2)
Wire length
Measuring rail
1)
1) 2)
A0043369
27