Proline Prosonic Flow P 500 Modbus RS485
Endress+Hauser
6.2.4
Mounting the sensor
WARNING
L
Risk of injury when mounting the sensors and strapping bands!
‣
Due to the increased risk of cuts, wear suitable gloves and protective goggles.
Sensor configuration and settings
DN 15 to 65 (½ to 2½")
Strapping band
2 traverses
[mm (in)]
1)
Sensor distance
Sensor distance
–
Wire length → 35
1)
Depends on the conditions at the measuring point (measuring pipe, medium etc.). The dimension can be
determined via FieldCare or Applicator. See also the Result sensor distance / measuring aid parameter in
the Measuring point submenu
2)
Only up to DN 600 (24")
Mounting types
Sensor holder with U-shaped screws
Can be used for
• Measuring devices with measuring range DN 15 to 65 (½ to 2½")
• Mounting on pipes DN 15 to 32 (½ to 1¼")
Procedure:
1. Disconnect the sensor from the sensor holder.
2. Position the sensor holder on the measuring pipe.
3. Fit the U-shaped screws through the sensor holder and lightly grease the thread.
4. Screw the nuts onto the U-shaped screws.
5. Position the sensor holder correctly and tighten the nuts uniformly.
15
Holder with U-shaped screws
1
Sensor holder
DN 50 to 4000 (2 to 160")
Strapping band
1 traverse
2 traverses
[mm (in)]
[mm (in)]
1)
Sensor distance
Measuring rail
1
Installation
Welded bolt
1 traverse
2 traverses
[mm (in)]
[mm (in)]
1)
1)
Sensor distance
Sensor distance
1) 2)
Wire length
Measuring rail
1)
1) 2)
A0043369
27