Commissioning
Parameter
Device tag
Parameter
Volume flow unit
Volume unit
28
8.3.1
Defining the tag name
To enable fast identification of the measuring point within the system, you can enter a
unique designation using the Device tag parameter and thus change the factory setting.
The number of characters displayed depends on the characters used.
Navigation
"Setup" menu → Device tag
Parameter overview with brief description
Description
Enter the name for the measuring point.
8.3.2
Setting the system units
In the System units submenu the units of all the measured values can be set.
Navigation
"Setup" menu → System units
System units
Volume flow unit
Volume unit
Parameter overview with brief description
Description
Select volume flow unit.
Result
The selected unit applies for:
• Output
• Low flow cut off
• Simulation process variable
Select volume unit.
Result
The selected unit is taken from: Volume
flow unit parameter
8.3.3
Configuring the pulse/frequency/switch output (option 3)
The Pulse/frequency/switch output 1 submenu contains all the parameters that must be
configured for the configuration of the selected output type.
Navigation
"Setup" menu → Pulse/frequency/switch output 1 to 2
User entry
Max. 32 characters, such as
letters, numbers or special
characters (e.g. @, %, /).
Selection
Unit choose list
Unit choose list
Dosimag
Factory setting
Dosimag
Factory setting
Country-specific:
• 138 = l/h
• 16 = gal/min (us)
Country-specific:
• 41 = l
• 40 = gal (us)
Endress+Hauser