Equipment Maintenance; General; Tank Draining - Omegasonics Viking Products Super Pro DX3 Operation & Instruction Manual

Ultrasonic cleaning equipment
Table of Contents

Advertisement

General

 Keep the bath free of oils, grease and any foreign materials.
 Skim off oil and grease residue periodically, if necessary.
 Cleaning agents should be changed periodically depending on usage.
Draining Tank
 Turn equipment off and unplug the power cord.
 Wait at least twenty (20) minutes after the heat circuit is turned off before emptying
the tanks. Permanent damage to the heater elements will occur if the tank is
drained too soon after the heaters are turned off.
 Drain the contaminated cleaning solutions from the tanks using all the valves.
 Rinse the inside of each tank with clean water.
 Buff the inside of the tanks with a clean, soft cloth. Do not use steel wool cleaning
pads as they are too abrasive and will scratch the tank surface.
 Rinse the tanks again.
 If the tanks will not be used for a long period of time, wipe the inside and the outside
of the tanks dry with a dry, clean, soft cloth.
 Close the lids on each tank. The lid should remain closed when the equipment is not
in use to keep dust and debris from accumulating.
 This tank cleaning procedure should be performed every time the bath is changed.
Always thoroughly inspect drain areas for leaks.
When discharging bath and waste, follow all environmental and regulatory requirements. A reputable and
licensed waste transportation firm should perform removal of all waste materials. Omegasonics is not
liable for improper handling of waste materials.

EQUIPMENT MAINTENANCE

22

Advertisement

Table of Contents
loading

This manual is also suitable for:

Omg-5038-dx3

Table of Contents