Digital Timer (Tank 1 and Tank 3)
Operation
 The 4-digit number on the upper
display is the amount of time that the
machine has been in actual
operation. It will count up to the
preset value.
 The illuminated 4-digit number on
the lower display is the preset value.
 Pressing the RESET button will
deactivate the timer and reset it to 0.
Turning the switch to RESET will
also deactivate the timer.
 Pressing the LOCK button will
prevent you from changing the
preset time value. LOCK will display
in the lower orange display.
 The ultrasound is factory set to 15:00
minutes.
 To adjust the preset value, press the
blue keys. Press the up arrow key, to
increase, or down arrow key, to
decrease.
 Left most blue key corresponds to 10
minute increments.
 Second blue key from the left
corresponds to 1 minute increments.
 Third blue key from the left
corresponds to 10 second
increments.
 Right most blue key corresponds to
1 second increments.
 The time is activated by turning the selector switched marked SONICS to the ON
position. The button will return to center. The "OP" light will illuminate on the lower,
left side orange display. A red light will illuminate and blink on the upper, left side red
display while the timer is operational.
15