Download Print this page

Biddle HR 12 Manual page 49

Advertisement

HR 12 M
ANUAL
5.2
Remedying errors that are accompanied by an error message
E
RROR MESSAGE
The malfunction lamp is lit up.
5.3
Reading out errors
5.4
Error codes registers
B
V
IT
ALUE
3
8
12
4096
E
F
[1];
RROR
LAGS
M
217
ODBUS
B
V
IT
ALUE
0
1
14
8192
Manual version 1.0 (15-07-2018)
P
ROBABLE CAUSE
The filter is letting insufficient air
through, due to contamination.
See also:
6.2
"Replacing the filter" on page 52
Errors are recorded with a binary code in two registers.
In the log file "log_error.csv", which can be read out with a
USB flash drive, the errors are indicated with a numeric value.
Error codes register 216/ErrorFlags[0]
_
.
LOG
ERROR
CSV
V
C
ALUE
ODE
3
E7
12
F8
Error codes register 217/ErrorFlags[1]
L
.
OGERROR
CSV
V
C
ALUE
ODE
16
F1
29
F17
W
HAT TO DO
Check the filter for contamination.
Clean or replace it if necessary.
You can adjust the pressure differ-
ence meter over the filter to make
it emit a dirty filter signal either
sooner or later.
D
ESCRIPTION
Fan error
Faulty pressure sensor in supply duct (X375)
D
ESCRIPTION
Filters must be replaced
Faulty CO
sensor
2
en
E
RRORS
en-49

Advertisement

loading

Related Products for Biddle HR 12