Summary of Contents for MSA Mini ID R-1234yf
Page 1
Mini ID R-1234yf REFRIGERANT IDENTIFIER OPERATION MANUAL Manual Part Number: 5-06-4900-75-0 Manual File:027245 Rev. B...
Page 2
Otherwise, it could fail to perform as designed, and persons who rely on this device could sustain serious injury or death. The warranties made by MSA with respect to the product are voided if the product is not installed and used in accordance with the instructions in this manual.
The warranties made by MSA with respect to the product are voided if the product is not used and serviced in accordance with the instructions in this manual. Please protect yourself and others by following them. We encourage our customers to write or call regarding this equipment prior to use or for any additional information relative to use or repairs.
2 Warnings and Cautions The Mini ID R-1234yf, hereafter also referred to as "the device," is a gas analyzer intended to provide a fast, easy, and accurate means to determine refrigerant purity in refrigerant storage cylinders or directly in air conditioning systems. The device is specified to support compliance with federal, state, and local safety codes that govern emissions.
Failure to follow these cautions can result in minor or moderate injury. 2.3 General Notices • Operate this device with vehicles or cylinders marked to contain R-1234yf. Cross contamination with other refrigerant types causes severe damage to the A/C system, to service tools, and equipment.
4 Product Description 4.1 General The Mini ID Refrigerant Identifier will provide an easy and accurate means to determine if the R-1234yf refrigerant in vehicle air conditioning systems is of suitable purity. The instrument utilizes non-dispersive infrared (NDIR) technology to determine the weight concentrations of R-1234yf.
5 Product Components 5.1 Base Module 5.2 Coupler with Filter 5.3 Battery Clips Manual Part Number: Manual File: P a g e 5-06-4900-75-0 027245 Rev. B...
6 Instruction for Use WARNING: Be sure to turn off the vehicle and let it rest for 3 minutes. Connect the power cable to the power port located on the bottom of the instrument. Connect the power clips to the 12 VDC battery source. Be sure to observe the correct polarity.
Page 11
If these steps fail to restore the unit to good working order, contact: MSA - The Safety Company 456 Creamery Way Exton, PA 19341 USA 1-610-524-8800 For your local MSA contacts, please go to our website www.MSAsafety.com Manual Part Number: Manual File: P a g e...