Table Of Contents; Software - National Instruments NI-9225 Calibration Procedure

3-channel, 300 vrms, 24-bit, simultaneous, channel-to-channel isolated analog input module
Table of Contents

Advertisement

CALIBRATION PROCEDURE
NI 9225
3-Channel, 300 Vrms, 24-Bit, Simultaneous,
Channel-to-Channel Isolated Analog Input Module
Français
This document contains the verification and adjustment procedures for the National Instruments
9225. For more information about calibration solutions, visit
Contents
Software.................................................................................................................................... 1
Documentation.......................................................................................................................... 2
Test Equipment......................................................................................................................... 3
Test Conditions......................................................................................................................... 3
Initial Setup............................................................................................................................... 4
Verification ............................................................................................................................... 4
Accuracy Verification....................................................................................................... 5
Gain Matching Verification.............................................................................................. 6
Phase Matching Verification ............................................................................................ 8
Adjustment................................................................................................................................ 9
Accuracy Adjustment ....................................................................................................... 9
EEPROM Update ..................................................................................................................... 10
Reverification ........................................................................................................................... 10
Accuracy Under Calibration Conditions .................................................................................. 10
Where to Go for Support .......................................................................................................... 11

Software

Calibrating the NI 9225 requires the installation of NI-DAQmx 9.3 or later on the calibration
system. You can download NI-DAQmx from
LabVIEW, LabWindows
need to install support for the application software that you intend to use.
Deutsch
ni.com/manuals
/CVI
, ANSI C, and .NET. When you install NI-DAQmx, you only
ni.com/calibration
ni.com/downloads
.
. NI-DAQmx supports