CALIBRATION PROCEDURE
NI 9229/9239
4-Channel, ±60 V/±10 V, 24-Bit Simultaneous, Channel-to-Channel
Isolated Analog Input Modules
This document contains the verification and adjustment procedures for the National
Instruments 9229/9239. For more information about calibration solutions, visit
.
calibration
Contents
Software.................................................................................................................................... 1
Documentation.......................................................................................................................... 2
Test Equipment......................................................................................................................... 3
Test Conditions......................................................................................................................... 3
Initial Setup............................................................................................................................... 3
Verification............................................................................................................................... 4
Accuracy Verification....................................................................................................... 4
Adjustment................................................................................................................................ 9
Accuracy Adjustment ....................................................................................................... 9
EEPROM Update ..................................................................................................................... 10
Reverification ........................................................................................................................... 10
Software
Calibrating the NI 9229/9239 requires the installation of NI-DAQmx 9.5 or later on the
calibration system. You can download NI-DAQmx from
supports LabVIEW, LabWindows
Basic .NET, and Microsoft Visual C#. When you install NI-DAQmx you only need to install
support for the application software that you intend to use.
™
/CVI
™
, ANSI C, Microsoft Visual C++, Microsoft Visual
ni.com/downloads
ni.com/
. NI-DAQmx
Need help?
Do you have a question about the NI 9229 and is the answer not in the manual?
Questions and answers