Kodak ThermoFlex Wide II Site Preparation Manual page 10

Output device
Table of Contents

Advertisement

6
Cables
Output device main power
Drum vacuum drive power
Workstation power
Chiller power
Debris removal cabinet power
Telephone line
a. The tolerances for input voltages are +6%, –10%. The specification for the frequency is 50/60 Hz
Branch circuits should have time delay type circuit breakers to avoid nuisance tripping caused by inrush
currents. For details, refer to Electrical on page 13.
Table 4: Cables and interconnects—Japan
Supplied by Kodak
Attached 4.5 m (14.7 ft.) power cord
with L6-30P plug
Labeled terminals for hookup of AC
mains input cable (3-phase plus
ground). Pilot hole for appropriate
conduit fitting.
2.5 m (8.2 ft.) power cable with 5-
15P connector
2.5 m (8.2 ft.) power cable with 5-
15P connector
See the UDRC Site Preparation Guide (part number 725-00089A).
Serial cable for modem.
ThermoFlex Wide II Site Preparation Guide
Supplied by customer
Dedicated 30 A, 200 V branch
circuit with slow-blow circuit
breaker and L6-30R receptacle
Dedicated 20A, 340-506 VAC
(±0%), 50/60 Hz, 3-phase branch
circuit with circuit breaker
protection.
Wiring, circuit breaker and panel
fittings supplied by customer in
accordance with local electrical
standards.
5-15R receptacle and dedicated
100 V, 15 A branch circuit
5-15R receptacle and dedicated
120 V, 15 A branch circuit
Two dedicated telephone lines and
RJ-11 cables to plug into modem
(must be long enough to reach
workstation). One line is to
connect to the workflow, and the
other is to connect to Kodak
remote support.
Also requires separate AC power
source (UPS provided only with 5-
15R receptacles).
a
±
1.5%.