Maintenance; Safety Instructions In Case Of Service Repairs - Orca UVpro EKB 100 Manual

Disinfection box
Table of Contents

Advertisement

Issue
There is condensation
between the UVC tube and
the jacket tube.
The disinfection effect
decreases and no longer
corresponds to the
interpretation.

MAINTENANCE

Safety instructions in case of service repairs

Remove the device from any power source before cleaning the tubes, or performing any repairs. The UVC
tubes should also be disconnected from power during visual inspections.
Do not, under any circumstances, perform a visual control on the UVC tubes without wearing eye-protection!
Protect your skin by wearing covering clothing, and never touch the UVC tubes without wearing gloves.
Before any repairs or work on the device, the device needs to be unplugged and
disconnected from any power source. Touch the glass tubes only while wearing gloves.
Eyes and skin must be protected from UVC radiation. Please always wear protective
goggles, gloves, and covering clothing while working in proximity to UVC tubes.
orca GmbH
Cause
The outer casing may be damaged.
If no cracks are visible, the water
must have penetrated through the
screw connection.
After 12,000 operating hours, the
UVpro tubes still provide about 75%
of the initial power. The plant is
configured for this value.
If the tubes have not yet operated
12,000 hours, they are possibly soiled
or calcified.
www.orcagmbh.de
Entkeimungsbox EKB 100
Fix
Check the glass for possible cracks.
Examine the area of the tube base
for wetness. Dry the socket and
replace the O-rings.
After commissioning, the tube
should be inspected promptly
(daily).
A tube replacement is
recommended.
Clean the tubes according to our
cleaning instructions.
Seite 15 von 20