Time control

The UVC disinfection box is to be switched on with the red rocker switch. With pushing the green start button,
the UV radiation starts for the pre-set time that was set by operating the rotary potentiometer. The red
flashing LED shows the course of time. Repeated pushing of the start button switches off the radiation
directly. To start the radiation, it is required that the door of the unit is closed.

Time setting

The radiation time is adjustable between ca. 1 to 32 min. (4 areas)
Normally, a radiation time of 5 min. is sufficient.
The time between 1 to 8 min can be read at the potentiometer. (area 1)

Pre-selection

The time setting can be programmed on 4 areas in total:
Area 1
1- 8 min.
Area 2
2- 16 min.
Area 3
3- 24 min.
Area 4
4- 32 min.
To change the area proceed as described:
1. Set rotary potentiometer to the correct value of the area (1-4).
2. Press green switch for a long time (ca. 3 sec.) until the red LED flashes.
3. The number of flashes shows the chosen area (can be still changed by the potentiometer).
4. Press green button again shortly. The pre-selection will be memorized.
Area 1 is pre-set in the delivery state and the pre-selection of the
potentiometer therefor is ca. 1-8 min.