Test Set I/O Command List - Keysight Technologies N52 Series User's And Service Manual

Millimeter head controller
Table of Contents

Advertisement

Service Information

Test Set I/O Command List

The following commands are available for controlling/reading the test set. The
step-by-step commands in the
procedure require the use of these commands. The user is required to select
the appropriate command, depending on the port or function under test.
Read commands involve a single number, an address. Write commands involve
two numbers, address and data.
Table 2-2
0
32,0
32,1
32,2
0,1
0,2
0,16
0,32
64,17
64,34
64,68
64,136
The amber Source LEDs are controlled separately from the switches for the
RF OUT paths.
The write commands for IF path control for the N5262A are listed in
on page
Table 2-5 on page
Keysight N5261/62A User's & Service Guide
I/O Commands
address for reading two digits from model number of the test set
turns on ALC for both SRC1 (Port 1 and 2) and SRC2 (Port 3 and 4)
turns off ALC for SRC1 (Port 1 and 2)
turns off ALC for SRC2 (Port 3 and 4)
SRC1 IN to Port 1 RF OUT
SRC1 IN to Port 2 RF OUT
SRC2 IN to Port 3 RF OUT
SRC2 IN to Port 4 RF OUT
turns on Receiver and Source LEDs for Port 1 (all other LEDs off)
turns on Receiver and Source LEDs for Port 2 (all other LEDs off)
turns on Receiver and Source LEDs for Port 3 (all other LEDs off)
turns on Receiver and Source LEDs for Port 4 (all other LEDs off)
76. The commands for reading and writing gain factors for are listed in
81.
Part Replacement and Location Information
Table 2-3
71