Commands

*OPC

Related Commands
Syntax

*OPT? (Query Only)

Related Commands
Syntax
Returns
Examples
3–26
The *OPC (Operation Complete) command generates the operation complete
message in the Standard Event Status Register (SESR) when all pending
operations finish. The *OPC? query places the ASCII character "1" into the
Output Queue when all pending operations are finished. The *OPC? response is
not available to read until all pending operations finish. For a complete
discussion of the use of these registers and the output queue, see page 3–35.
*WAI, ADJusting?
The *OPT (Option Identification Query) command returns a string that identifies
what options are installed in the VXOA41.
N/A
The command returns a single " " if the VXOA41 is in standard configuration
(no options, single-mode fiber). If options are installed, the command returns a
string of three fields, with each field separated by a comma. Each of the fields
represents one of the possible options. Since the options are mutually exclusive,
two of the fields will always be "0". One of the fields, which one depends on
what option is installed, will indicate the installed option with the syntax:
"
"
Might return the string:
VXOA41 Optical Attenuator User Manual