Selecting the Cephalometric Blade Collimator
Parameters
To select the cephalometric blade collimator parameters, follow these steps
1. From the fixed arm, on the control panel, on
same time, until you see Collimator Calibration Parameters displayed on the Display
Screen of the control panel.
2. Press
to enter the sub-menus. The following sub-menus are displayed:
•
Pano: Panoramic parameters
•
Ceph: Cephalometric parameters
•
3D: 3D parameters
3. Press
to select Ceph. The following parameters are displayed:
•
ORG: ORG X and ORG Y
•
DIM: DIM X and DIM Y
Facing the digital sensor:
•
A is the left blade
•
B is the right balde
•
C is the top blade
•
D is the bottom balde
+
Dim Y
A
+
Org Y
C
+
Org X
D
KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed 01
press and hold
and
B
+
Dim X
Post-Installation Control
at the