Chapter 6
Maintenance
Annual Maintenance
We recommend that a general inspection of the unit should be carried out every year by an
approved dental systems technician.
The inspection should cover the following points:
•
Check the attachment points to the floor and the wall.
•
Check all the mobile components of the unit.
•
Check the x-ray generator.
•
Check the alignment of the x-ray.
•
Make an image acquisition with the test tools and check the image.
•
Check the focal trough and the symmetry.
•
Check the degree of legibility of the labels.
•
Check for damage to cables, covers, oil leaks, etc.
If the results of any of these inspections are unsatisfactory, refer to the Service Manual in order
to rectify any problems.
If there is any doubt, do not operate the unit.
KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed 01
6–1