3. OPERATION

Leveling the equipment
The MB27 has a level bubble in a small round window near the display.
Make sure that the bubble is centered in the circle and that he equipment is level each time its location is changed.
Display
Controls
Button:
Name:
Start / Stop
Tare
Temp
Time
There are also several modes of operation which affect how each button operates. Please refer to the following section.
Operating Modes
Standby Mode
When the Analyzer is connected to AC power and the display is off, the unit is in Standby mode.
Start
(Short press) Turns on the display and enters Weigh mode.
Print
(Long press) Displays the RS232 settings (see Section 6, RS232 Data Connection.). Thereafter,
Set 
(Short or long press) Switch the RS232 setting between two settings. (see section 6, RS232 Data Settings)
Set 
(Short or long press) Switch the RS232 setting between two settings. (see section 6, RS232 Data Settings)
Print
(Short press) Re-enters Standby mode (off).
Weigh Mode
The display shows the weight of items placed on the pan support.
Tare
(Short press) Sets the displayed weight value to zero
Start
(Long press) Enters Standby mode (Off)
The following operations can be initiated:
Temp
(Short press) Enters Edit mode for the temperature setting, the value will blink
Time
(Short press) Enters Edit mode for the time setting, the value will blink
%
%
g
100°
10:00
Quick Functions:
On (short press) / Off (long press)
Start/Stop (short press)
Back (short press)
Tare (short press)
Enter / Accept value (short press)
Temperature Setting (short press)
Time Setting (short press)
Stable indicator
Percent solids
Percent moisture
grams
Temperature setting or current chamber temperature (Celsius)
Time setting (minutes : seconds)
Button:
Name:
Quick Functions:
Increase value
Set 
(short or long press)
Decrease value
Set 
(short or long press)
Print (short press)
Print / Cal
Calibrate (long press)
%g
Result Unit (short press)