iseg CalPower EHQ 102M Operator's Manual

iseg CalPower EHQ 102M Operator's Manual

Precision high voltage power supplies in 3u eurocard format with usb and rs232-interface

Advertisement

Quick Links

Via Acquanera, 29
22100 Como
tel. 031.526.566 (r.a.) fax 031.507.984
info@calpower.it
www.calpower.it
Precision High Voltage Power Supplies in 3U Eurocard Format
Contents:
1. General information
2. Technical Data
3. EHQ Description
4. Front panel
5. Handling
6. RS232 Interface
7. Program example
Appendix A: Block diagram
Appendix B: Rotary switch locations
Appendix C: USB Interface including
Attention!
-It is not allowed to use the unit if the covers have been removed.
-We decline all responsibility for damages and injuries caused by an improper use of the mo-
dule. It is highly recommended to read the operators manual before any kind of operation.
Note
The information in this manual is subject to change without notice. We take no responsibility whatsoever for any
error in the document. We reserve the right to make changes in the product design without reservation and
without notification to the users.
Filename EHQ10x_02_eng.___; Version 3.14 from 2013-09-04
iseg Spezialelektronik GmbH
Bautzner Landstr. 23
D - 01454 Radeberg / Rossendorf
EHQ 102M / 103M / 104M / 105M
with USB and RS232-Interface
Operators Manual
new command list
Email: sales@iseg-hv.de
http://www.iseg-hv.com
Germany
S p e z i a l e l e k t r o n i k G m b H
Phone ++ 49 351 / 26 996 - 0
Fax
++ 49 351 / 26 996 - 21

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CalPower EHQ 102M and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for iseg CalPower EHQ 102M

  • Page 1 We reserve the right to make changes in the product design without reservation and without notification to the users. Filename EHQ10x_02_eng.___; Version 3.14 from 2013-09-04 iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 996 - 0 Bautzner Landstr.
  • Page 2 S p e z i a l e l e k t r o n i k G m b H iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 996 - 0 Bautzner Landstr. 23 http://www.iseg-hv.com ++ 49 351 / 26 996 - 21...
  • Page 3: General Information

    USB mini B on the front panel HV connector SHV-Connector at the front panel Inhibit connector 1-pin Lemo-hub iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 996 - 0 Bautzner Landstr. 23 http://www.iseg-hv.com ++ 49 351 / 26 996 - 21...
  • Page 4: High Voltage Supply

    The high-voltage filters are housed in individual metal enclosures to shield even minimal interference radiation. iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 996 - 0 Bautzner Landstr.
  • Page 5: Front Panel

    If the function ″Autostart″ has been activated in the previous operating session, the high Attention! voltage generation starts immediately with the saved parameters! iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 996 - 0 Bautzner Landstr. 23 http://www.iseg-hv.com...
  • Page 6: Manual Control Mode

    While the unit is operated in manual control mode, RS232 read cycles are interpreted only. Commands are accepted, but do not result in a change of the output voltage. iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 996 - 0 Bautzner Landstr.
  • Page 7 <CR> <LF> ( 0x0D 0x0A , 13 10 respectively). Leading zeroes can be omitted on input, output is in fixed format. iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 996 - 0 Bautzner Landstr. 23 http://www.iseg-hv.com...
  • Page 8: Command Set

    If output voltage has been shut off permanently (by ERR or INH at ENABLE KILL or TRP) the command ”Read status word” must be executed before the output voltage can be restored. iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 996 - 0 Bautzner Landstr.
  • Page 9: Error Codes

    (EEPROM guarantees a minimum of 1 million saving cycles) Software Contact us for an overview on our user friendly control and data acquisition software! iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 996 - 0 Bautzner Landstr. 23 http://www.iseg-hv.com...
  • Page 10: Program Example

    S p e z i a l e l e k t r o n i k G m b H 7. Program example /***************************************************************************************************/ ehq.cpp example program for iseg ehq hv boards, written by Jens Römer, 27.2.97 this code was compiled under BC, please contact iseg for the source file /***************************************************************************************************/ #include <dos.h> #include <stdio.h> #include <conio.h>...
  • Page 11 HV-Filter OUTPUT switch HV-Transformer Appendix A: Block diagram EHQ iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 996 - 0 Bautzner Landstr. 23 http://www.iseg-hv.com ++ 49 351 / 26 996 - 21 D - 01454 Radeberg / Rossendorf...
  • Page 12 Appendix B: EHQ side cover Polarity rotary switch (e.g.: polarity negative) Rotary switches for V and I iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 996 - 0 Bautzner Landstr. 23 http://www.iseg-hv.com ++ 49 351 / 26 996 - 21...
  • Page 13 EDCP *INSTR,SCPI *INSTR,EDCP select SCPI instruction set *INSTR,iseg *INSTR,DCP select old iseg instruction set iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 996 - 0 Bautzner Landstr. 23 http://www.iseg-hv.com ++ 49 351 / 26 996 - 21...
  • Page 14 S p e z i a l e l e k t r o n i k G m b H iseg SCPI command set Command Description :VOLTage set channel voltage <Voltage>[V] shut channel emergency off <EMCY OFF> clear shut channel emergency off <EMCY CLR>...
  • Page 15 CONF:ECHO ON Advanced Examples: :VOLT 2000.5; :READ:VOLT?; :CURR 0.002; :READ:CURR? :MEAS:VOLT?; CURR? 20000.284V; 1999.731E-6A iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 996 - 0 Bautzner Landstr. 23 http://www.iseg-hv.com ++ 49 351 / 26 996 - 21...
  • Page 16 (OPTION) iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 996 - 0 Bautzner Landstr. 23 http://www.iseg-hv.com ++ 49 351 / 26 996 - 21...
  • Page 17 (If Vset has been set to a value unequal to zero (0V) before the status bit ‘isOn’ is changed from (1) one to (0) zero a ramp down of the voltage to zero (0V) will be started.) iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 996 - 0 Bautzner Landstr.
  • Page 18 Event: A hardware failure of the HV module has been detected. The HV is ESrvs EventService switched off without the possibility to switch on again. Please consult the iseg Spzialelektronik GmbH. Reserved iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 996 - 0 Bautzner Landstr.
  • Page 19 Hardware ClearKill signal and clear all event signals of the module and the channels doCLEAR=0 no action Common-instructions The common instruction set is in depended from the selected language (iseg or SCPI) and can be used always. query module identification *IDN? reply iseg Spezialelektronik GmbH,EHQ 103,480403,3.00...

This manual is also suitable for:

Calpower ehq 103mCalpower ehq 104mCalpower ehq 105m

Table of Contents