16-channel high voltage bipolar power supply (4 pages)
Summary of Contents for iseg HPS Series
Page 1
We reserve the right to make changes in the product design without any notification to the users. Filename HPx_300-800W_03_eng.___; Version 4.04 as of 2006-11-08 iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Tel ++ 49 (0)351 / 26 996 - 0 Bautzner Landstr.
Page 2
S p e z i a l e l e k t r o n i k G m b H iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Tel ++ 49 (0)351 / 26 996 - 0 Bautzner Landstr. 23 http://www.iseg-hv.de Fax ++ 49 (0)351 / 26 996 - 21...
(> 15 s) of the output capacitance has to be kept and the discharge status has to be checked afterwards. Only skilled and authorised people are allowed to do any service, repair or maintenance for this unit. iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Tel ++ 49 (0) 351 / 26 996 - 0 Bautzner Landstr.
Page 4
The output voltage will be shut off permanently without ramp, if I OUTmax The re-setting of the output voltage is possible after pushing button “HV-ON“ again. iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Tel ++ 49 (0)351 / 26 996 - 0 Bautzner Landstr.
Page 5
If no push button of the menu will be switched the display is automatically going back to the mode „HV-OFF“ after a delay time of 30 sec. iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Tel ++ 49 (0) 351 / 26 996 - 0 Bautzner Landstr.
Page 6
On ⇒“Set”⇒Off⇒“Set” ⇒On Echo on/off for RS232 F 13 ECHO “SET“ “ENTER“ Back to F 01 iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Tel ++ 49 (0)351 / 26 996 - 0 Bautzner Landstr. 23 http://www.iseg-hv.de Fax ++ 49 (0)351 / 26 996 - 21...
Environment conditions Operating temperature: 5 up to 35 °C Humidity: 30% up to 80 %, no condensation iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Tel ++ 49 (0) 351 / 26 996 - 0 Bautzner Landstr. 23 http://www.iseg-hv.de Fax ++ 49 (0) 351 / 26 996 - 21...
Page 8
S p e z i a l e l e k t r o n i k G m b H Storage temperature 0 up to 60 °C iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Tel ++ 49 (0)351 / 26 996 - 0 Bautzner Landstr.
Page 9
5 up to 35 °C Humidity: 30% up to 80 %, no condensation Storage temperature 0 up to 60 °C iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Tel ++ 49 (0) 351 / 26 996 - 0 Bautzner Landstr. 23 http://www.iseg-hv.de Fax ++ 49 (0) 351 / 26 996 - 21...
Page 10
S p e z i a l e l e k t r o n i k G m b H 4. Description of the CAN Interface. The integrated CAN-Interface of the HPS series offers 2 functions. Either the interface gives access to the implemented firmware of the processor controlled unit, this is...
Page 11
2 3 4 5 6 7 8 Message with module 7E7h address and the according (RTR=0) ID´s iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Tel ++ 49 (0) 351 / 26 996 - 0 Bautzner Landstr. 23 http://www.iseg-hv.de Fax ++ 49 (0) 351 / 26 996 - 21...
Page 13
The transfer of commands works in ASCII-code. The end of a set of characters is made by <CR><LF> ($0D $0A or 13 10). On input side the leading zeros are useless, the output side is in fixed format without leading zero. iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Tel ++ 49 (0) 351 / 26 996 - 0 Bautzner Landstr.
Page 14
IL,100mA Command to set the ramp speed of the output voltage: RAMP,<rampspeed>V/s example RAMP,1000V/s iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Tel ++ 49 (0)351 / 26 996 - 0 Bautzner Landstr. 23 http://www.iseg-hv.de Fax ++ 49 (0)351 / 26 996 - 21...
Page 16
(no Look At Me status has been found ) Read of unit identifier response example ID, iseg Spezialelektronik r3.02 sn.680041 Type HPN 30 107 iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Tel ++ 49 (0)351 / 26 996 - 0 Bautzner Landstr. 23 http://www.iseg-hv.de...
Page 17
Response example Ramp, RANGE=3000 V/s, VALUE=1000 V/s Read out of unit identifier: :READ:IDNT? Response example ID, iseg Spezialelektronik r3.02 sn.680043 Type HPN 30 107 iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Tel ++ 49 (0) 351 / 26 996 - 0 Bautzner Landstr. 23 http://www.iseg-hv.de...
Page 18
LAM,OK (no Look At Me status has been found) iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Tel ++ 49 (0)351 / 26 996 - 0 Bautzner Landstr. 23 http://www.iseg-hv.de Fax ++ 49 (0)351 / 26 996 - 21...
Page 19
S p e z i a l e l e k t r o n i k G m b H 7.3 Common commands *RST reset the device *IDN? Response example ID, iseg Spezialelektronik r3.02 sn.680034 Type HPN 30 107 *GTL go to local, local-button is enabled *LLO local logout, Local-Button is disabled *CLS clear Status Software release 4.02:...
Page 20
37,5 mA output current 1.0 V control voltage corresponds to 15 mA output current iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Tel ++ 49 (0)351 / 26 996 - 0 Bautzner Landstr. 23 http://www.iseg-hv.de Fax ++ 49 (0)351 / 26 996 - 21...
Page 21
If these provisions do not lead to a good result this unit has to checked from an authorised agent or must be shipped to the factory. iseg Spezialelektronik GmbH Email: sales@iseg-hv.de Tel ++ 49 (0) 351 / 26 996 - 0 Bautzner Landstr.
Need help?
Do you have a question about the HPS Series and is the answer not in the manual?
Questions and answers