Page 1
S p e z i a l e l e k t r o n i k G m b H 9-channel High Voltage Power Supply EHQ 9005-F Operators Manual Contents 1. General information 2. Technical data 3. Handling 4. Communication via interface 4.1 Device Protocol DCP...
Page 2
S p e z i a l e l e k t r o n i k G m b H iseg c/o ROTECH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 95 - 260 Spezialelektronik Bautzner Landstr. 45 http://www.iseg-hv.com...
The outputs (V-) und (V+) of each channel are both floating against each other and against ground. The EHQ 9005-F is made ready for mounting into a crate. It is also possible to supply the modules separately with the necessary power. The unit is software controlled via CAN Interface directly through a PC or similar controller.
Page 4
The supply voltages and the CAN interface is connected to the module via a 96-pin connector on the rear side of the module. The 9-channel Module EHQ 9005-F is assembled of two sub-modules (8 channels / 1 channel), each controlled independently via an own CAN identifier.
Page 5
Transport Layer and functions of the Data Link Layer may be possible. The communication between the controller and a module on the same bus segment will be described as follows. iseg c/o ROTECH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 95 - 260 Spezialelektronik Bautzner Landstr. 45 http://www.iseg-hv.com...
Page 6
S p e z i a l e l e k t r o n i k G m b H 4.2 Summary of CAN data frames The 9-channel Module EHQ 9005-F is assembled of two sub-modules (8 channels / 1 channel), each controlled independently via an own CAN identifier.
Page 7
G1 to G0: Group 0 to 3 Module is log-off Only needed if group controller (GC) is used iseg c/o ROTECH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 95 - 260 Spezialelektronik Bautzner Landstr. 45 http://www.iseg-hv.com ++ 49 351 / 26 95 - 261 GmbH...
Page 8
With help of the ‘Group access’ ‘Switch ON /OFF’ the concerning channels are switched ON again. Function will be switched off with write ‘Current trip = 0’. iseg c/o ROTECH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 95 - 260 Spezialelektronik Bautzner Landstr. 45 http://www.iseg-hv.com...
Page 9
Actual current with resolution I / 10∗exp6 [A] O max in DATA_2 to DATA_0 iseg c/o ROTECH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 95 - 260 Spezialelektronik Bautzner Landstr. 45 http://www.iseg-hv.com ++ 49 351 / 26 95 - 261...
Page 10
Set voltages higher than the maximum module voltage will be ignored and the bit ‘Input error’ of the ‘Status channel’ will be set. iseg c/o ROTECH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 95 - 260 Spezialelektronik Bautzner Landstr. 45 http://www.iseg-hv.com...
Page 11
⇒ V shut of 0 V because software current trip has been exceeded iseg c/o ROTECH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 95 - 260 Spezialelektronik Bautzner Landstr. 45 http://www.iseg-hv.com ++ 49 351 / 26 95 - 261...
Page 14
If the averaging is ‘ON’ then the voltage and current measurement will work with a ‘Weighted Average Calculation’ of 16 ADC measurement values. iseg c/o ROTECH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 95 - 260 Spezialelektronik Bautzner Landstr. 45 http://www.iseg-hv.com...
Page 15
(ID9 = 0) than normal messages, only if one of the sumstatus- and safety loop-bits in the group access “General status module” not has been set. iseg c/o ROTECH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 95 - 260 Spezialelektronik Bautzner Landstr. 45 http://www.iseg-hv.com...
Page 16
0 V ramping to set voltage will be started automatically again. The green LED flashes on again only after the Group access ‘Status2 Current limit’ has been read. iseg c/o ROTECH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 95 - 260 Spezialelektronik Bautzner Landstr. 45 http://www.iseg-hv.com...
Page 18
Set voltages higher than the maximum specific module voltage are ignored and the bit ‘Input Error’ in ‘Status channel’ will be set. iseg c/o ROTECH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 95 - 260 Spezialelektronik Bautzner Landstr. 45 http://www.iseg-hv.com...
Page 19
- invalid bit rates will be ignored from the module and the bit ‘Input error’ of the ‘Status channel 0’ will be set. iseg c/o ROTECH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 95 - 260 Spezialelektronik Bautzner Landstr. 45 http://www.iseg-hv.com...
Page 20
DATA_0 Designation DATA _DIR Data Description write x = 2: with iseg Standard-CAN message ID9 is always dominant x = 4: with iseg Active-CAN message ID9 is recessive iseg c/o ROTECH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 95 - 260 Spezialelektronik Bautzner Landstr.
Page 23
With help of the ‘Group access’ ‘Switch ON /OFF’ the concerning channels are switched ‘ON’ again. iseg c/o ROTECH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 95 - 260 Spezialelektronik Bautzner Landstr. 45 http://www.iseg-hv.com...
Page 24
In one CAN segment only modules are allowed with different identifiers and the same bit rates. The factory fixed bit rate is written on the sticker of the 96-pin connector. iseg c/o ROTECH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 95 - 260 Spezialelektronik Bautzner Landstr. 45 http://www.iseg-hv.com...
Page 25
“General status module” has not been set (the module must be configured as a CAN-node with an Active-CAN message function). iseg c/o ROTECH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 95 - 260 Spezialelektronik Bautzner Landstr. 45 http://www.iseg-hv.com...
S p e z i a l e l e k t r o n i k G m b H Appendix A: Side view Desk open, jumper for safety-loop iseg c/o ROTECH Email: sales@iseg-hv.de Phone ++ 49 351 / 26 95 - 260 Spezialelektronik Bautzner Landstr. 45 http://www.iseg-hv.com...