Test Procedure CIC

Doc 0452940 rev. A
1- Open Test Control. (DSA 6000)
2- Ensure Acoustical Calibration and Validation System
(Doc# 0202830) have been calibrated prior to testing
hearing instruments.
3- Wrap the canal tip with Fun-tak including the vent hole.
4- Leave vent on faceplate side open.
5- Check to see that the receiver tube is centered in the
coupler and the tip of the canal is flush with the coupler.
6- Connect programming cable and battery pill into device.
7- Place the device in the sound chamber box making sure
that it is on the blue area with the faceplate facing the
speaker.
Close the sound chamber and run the test.
Follow instructions on screen.
Page 14 of 14