Page 2
My hearing aid information (if se Product: ReSound ONE Model: RT61-DRW Serial number (R): _______________________________ Serial number (L): _______________________________ Battery size: 312...
Page 3
Programs: (if set up by your hearing care professional) PROGRAM WHEN TO USE BEEPS...
Page 4
Get to know your hearing aid A. Program button B. Microphones C. Receiver D. Dome View ReSound’s hearing aid tutorial videos...
Page 5
E. Battery door F. Right/Left indicator G. Serial number and model rial videos at youtube.com/resoundus.
What you need to know Getting started: On/Off function ON: Insert battery & completely close battery door (E). Program 1 will be active. OFF: Open the battery door (E). Use fingertip to pull it all the way.
Inserting/Replacing the battery: 1. Open the battery door (E) completely by using your fingertip. 2. Remove protective sticker from battery and wait two minutes before inserting the battery into the hearing aid to allow full activation of the battery.
Page 8
Flig 3. Insert the new battery of w with the positive side (flat side) tran in the correct position, as shown. by y 4. Gently close the battery door. If it doesn’t close easily, check to see if the battery is upside down. flig...
Page 9
Flight mode: Some airlines prohibit the use of wireless hearing aids (RF transmitters) onboard. If required by your airline, enable flight mode: Disable wireless operation by opening and closing the battery door three times within ten seconds (open-close, open-close, open-close). Your instruments will now be in flight mode.
Page 10
For more information Refer to the user guide included in your hearing aid box. Your hearing care professional can be reached at: View ReSound hearing aid tutorial videos at youtube.com/resoundus. Manufacturer according to FDA: GN ReSound North America 8001 E. Bloomington Freeway...
Page 11
Do not use water when you are cleaning the receiver tubes. As needed: Change domes (D) ReSound standard domes 1. Push the new dome over the ribbed end of the receiver wire. 20.07...
Page 12
When placed correctly, the dome will appear slightly angled. ReSound tulip domes Tulip domes consist of two “petals”. It is important that the largest petal is the outermost petal. To ensure this: 1.
Page 13
2. In Monthly: Change the wax guard or see your hearing care professional. For changing white wax guards, complete the following steps: 1. Removal: Remove the dome from the receiver wire. Open the wax filter case and take one of the tools out. s”.
Page 14
Ding (5 or 10) Hearing aid is starting u lter t is Ding-ding, Hearing aid is in flight m t on ding-ding, etc. Ripple tone Hearing aid is in stream mode View ReSound hearing aid tutorial videos at youtu...
Page 15
the sound outlet until the outer ring is touching the sound outlet. Pull the tool straight out – your new wax filter will remain in place. Re-attach your dome or a replacement dome. ACTION REQUIRED Change battery is starting up After dings cease, hearing aid is turned on is in flight mode...
Troubleshooting SYMPTOM CAUSE No sound Battery door (E) not c Dead battery Battery door (E) will n Blocked earmold or tu Blocked wax guard Not loud enough Blocked earmold or do Change in hearing, vo Excessive ear wax Excessive Incorrect earmold pla whistling/feedback Incorrect dome (D) pl Excessive ear wax...
Page 17
Re-insert dome (D) ear wax Clean guard/consult your hear the above Consult your hearing care prof y be in flight mode Open and close the battery Wireless operation will reactiv mentioned in this guide, please contact ReSound or your h...
Page 18
the battery door (E) erly or consult your hearing care professional ace dome (D), replace guard ng care professional ng care professional carefully lt your hearing care professional ng care professional door e battery (E) once. will reactivate ten seconds later. nd or your hearing care professional.
Need help?
Do you have a question about the ONE 61 RIE and is the answer not in the manual?