AD water
connection for 30
- 1000 kPa
pressure -
pressure-
resistant
(optional)
– Install the filter (supplied in accessory pack) between the stopcock
valve and the inlet hose. The filter for AD water is made of
chromium-nickel steel and can be recognised by its dull surface.
IMPORTANT
Australia and New Zealand
For Australia and New Zealand a non-return valve is required
between the tap and filter.
Do NOT shorten or otherwise damage the inlet hoses (see
diagram).
See installation diagram supplied.
This machine can be optionally supplied for a pressurised system
operating between 30 - 1000 kPa. If the water pressure is below
200 kPa the water intake duration will be automatically increased.
The pressure tested hose for AD water, coded green, has a ¾ inch
threaded union for connection to the onsite stopcock for AD water.
If the machine is not going to be connected to AD water, the AD
water connection has to be deactivated by a Miele Service
technician. The inlet hose remains in position at the back of the
machine.