Designates a possible imminent risk. If this is not avoided, it could lead to minor injuries. NOTICE NOTICE Designates a possibly harmful situation. If this is not avoided, the products could be damaged. Instructions for use • Micromanipulator Interface VITOM 3D • MIC912_EN_V1.0_07-2021_IFU_CE-MDR...
Normal use 2 Normal use 2.1 Intended use Micromanipulator interfaces are intended to attach laser micromanipulators to VITOM 3D exoscopes during microsurgical, open and endoscopic procedures. Micromanipulator interfaces do not have body contact. 2.2 Indications Indications for the micromanipulator interfaces result from the use of VITOM exoscope.
Products that come into contact with the central nervous system can become contaminated by organic residue containing prions. Prions lead to infection with Creutzfeldt-Jakob disease. If Creutzfeldt-Jakob disease has been diagnosed or is suspected: Do not continue to use the product. Instructions for use • Micromanipulator Interface VITOM 3D • MIC912_EN_V1.0_07-2021_IFU_CE-MDR...
Page 7
Safety Dispose of the product properly. Instructions for use • Micromanipulator Interface VITOM 3D • MIC912_EN_V1.0_07-2021_IFU_CE-MDR...
Base plate with thread for the dovetail connection adaptor from Leica, Zeiss / Möller-Wedel Rotating wheel for securing the height adjustment Screw for fastening the clamping mechanism to the VITOM 3D TH004 mounted on the VITOM 3D 4.2 Possible combinations It is recommended that the suitability of the products for the planned procedure be checked prior to use.
A code number after the CE mark indicates the responsible notified body. 4.4 Ambient conditions There are no special transport and storage conditions for this product. Instructions for use • Micromanipulator Interface VITOM 3D • MIC912_EN_V1.0_07-2021_IFU_CE-MDR...
When changing loads, it may be necessary to support the holding arm in order to prevent uncontrolled movement. Attach the clamping mechanism of the micromanipulator interface to the sheath of the VITOM 3D and lock in place by rotating the screw. Instructions for use • Micromanipulator Interface VITOM 3D • MIC912_EN_V1.0_07-2021_IFU_CE-MDR...
Page 11
The height of the adaptor can also be set using the rotating wheel for height adjustment (see arrow). The dovetail connection adaptor of the microscope model being used and the micromanipulator can now be mounted and aligned on the adaptor base plate. Instructions for use • Micromanipulator Interface VITOM 3D • MIC912_EN_V1.0_07-2021_IFU_CE-MDR...
Page 12
VITOM 3D. When using the UR300, the VITOM 3D must be not moved into a horizontal position. To prevent interference, the 3D glasses must be worn on top of the laser goggles for laser applications.
Disassembly 6 Disassembly 6.1 Removing the TH004 micromanipulator interface Unscrew the screw for fastening the clamping mechanism. Open the clamping mechanism and remove it from the VITOM 3D sheath. Instructions for use • Micromanipulator Interface VITOM 3D • MIC912_EN_V1.0_07-2021_IFU_CE-MDR...
7 Maintenance, servicing, repairs, and disposal 7.1 Repairing the product Repair work may only be performed by KARL STORZ or by a company authorized by KARL STORZ. The interventions described in this instruction manual are exempt from this rule. Please contact your local KARL STORZ subsidiary or authorized dealer (see the list of subsidiaries).