Instructions for use
Micromanipulator Interface VITOM 3D

Advertisement

Table of Contents
loading

Summary of Contents for Storz VITOM 3D

  • Page 1 Instructions for use Micromanipulator Interface VITOM 3D...
  • Page 2 07-2021 Copyright © All product illustrations, product descriptions, and texts are the intellectual property of KARL STORZ SE & Co. KG. Their use and reproduction by third parties require the express approval of KARL STORZ SE & Co. KG. All rights reserved.
  • Page 3: Table Of Contents

    5 Preparation  ..........................10 5.1 Unpacking the product ....................10 5.2 Inspecting the device  ...................... 10 5.3 Fastening the micromanipulator interface TH004 onto the VITOM 3D sheath ....10 6 Disassembly  ..........................13 6.1 Removing the TH004 micromanipulator interface ............13 7 Maintenance, servicing, repairs, and disposal  ................. 14 7.1 Repairing the product ......................
  • Page 4: General Information

    Designates a possible imminent risk. If this is not avoided, it could lead to minor injuries.   NOTICE    NOTICE Designates a possibly harmful situation. If this is not avoided, the products could be damaged. Instructions for use • Micromanipulator Interface VITOM 3D • MIC912_EN_V1.0_07-2021_IFU_CE-MDR...
  • Page 5: Normal Use

    Normal use 2 Normal use 2.1 Intended use Micromanipulator interfaces are intended to attach laser micromanipulators to VITOM 3D exoscopes during microsurgical, open and endoscopic procedures. Micromanipulator interfaces do not have body contact. 2.2 Indications Indications for the micromanipulator interfaces result from the use of VITOM exoscope.
  • Page 6: Safety

    Products that come into contact with the central nervous system can become contaminated by organic residue containing prions. Prions lead to infection with Creutzfeldt-Jakob disease. If Creutzfeldt-Jakob disease has been diagnosed or is suspected: Do not continue to use the product. Instructions for use • Micromanipulator Interface VITOM 3D • MIC912_EN_V1.0_07-2021_IFU_CE-MDR...
  • Page 7 Safety Dispose of the product properly. Instructions for use • Micromanipulator Interface VITOM 3D • MIC912_EN_V1.0_07-2021_IFU_CE-MDR...
  • Page 8: Product Description

    Base plate with thread for the dovetail connection adaptor from Leica, Zeiss / Möller-Wedel Rotating wheel for securing the height adjustment Screw for fastening the clamping mechanism to the VITOM 3D  TH004 mounted on the VITOM 3D 4.2 Possible combinations It is recommended that the suitability of the products for the planned procedure be checked prior to use.
  • Page 9: Symbols On The Packaging

    A code number after the CE mark indicates the responsible notified body. 4.4 Ambient conditions There are no special transport and storage conditions for this product. Instructions for use • Micromanipulator Interface VITOM 3D • MIC912_EN_V1.0_07-2021_IFU_CE-MDR...
  • Page 10: Preparation

    When changing loads, it may be necessary to support the holding arm in order to prevent uncontrolled movement. Attach the clamping mechanism of the micromanipulator interface to the sheath of the VITOM 3D and lock in place by rotating the screw. Instructions for use • Micromanipulator Interface VITOM 3D • MIC912_EN_V1.0_07-2021_IFU_CE-MDR...
  • Page 11 The height of the adaptor can also be set using the rotating wheel for height adjustment (see arrow). The dovetail connection adaptor of the microscope model being used and the micromanipulator can now be mounted and aligned on the adaptor base plate. Instructions for use • Micromanipulator Interface VITOM 3D • MIC912_EN_V1.0_07-2021_IFU_CE-MDR...
  • Page 12 VITOM 3D. When using the UR300, the VITOM 3D must be not moved into a horizontal position. To prevent interference, the 3D glasses must be worn on top of the laser goggles for laser applications.
  • Page 13: Disassembly

    Disassembly 6 Disassembly 6.1 Removing the TH004 micromanipulator interface Unscrew the screw for fastening the clamping mechanism. Open the clamping mechanism and remove it from the VITOM 3D sheath. Instructions for use • Micromanipulator Interface VITOM 3D • MIC912_EN_V1.0_07-2021_IFU_CE-MDR...
  • Page 14: Maintenance, Servicing, Repairs, And Disposal

    7 Maintenance, servicing, repairs, and disposal 7.1 Repairing the product Repair work may only be performed by KARL STORZ or by a company authorized by KARL STORZ. The interventions described in this instruction manual are exempt from this rule. Please contact your local KARL STORZ subsidiary or authorized dealer (see the list of subsidiaries).
  • Page 15: Subsidiaries

    Phone: +358 (0)96824774, Fax: +358 (0)968247755 E-mail: asiakaspalvelu@karlstorz.fi KARL STORZ Endoskope Greece M.E.P.E.* Patriarhou Grigoriou E’ 34, 54248 Thessaloniki, Greece Phone: +30 2310 304868, Fax: +30 2310 304862 E-mail: info-gr@karlstorz.com *Repair & Service Subsidiary Instructions for use • Micromanipulator Interface VITOM 3D • MIC912_EN_V1.0_07-2021_IFU_CE-MDR...
  • Page 16 Room 701A Building 5 & Room 501 Building 7, No. 3000 Longdong Avenue, Pilot Free Trade Zone, 201203, Shanghai, People’s Republic of China Phone: +86 21 60339888, Fax: +86 21 60339808 E-mail: info@karlstorz.com.cn Instructions for use • Micromanipulator Interface VITOM 3D • MIC912_EN_V1.0_07-2021_IFU_CE-MDR...
  • Page 17 Subsidiaries Instructions for use • Micromanipulator Interface VITOM 3D • MIC912_EN_V1.0_07-2021_IFU_CE-MDR...
  • Page 18 Subsidiaries Instructions for use • Micromanipulator Interface VITOM 3D • MIC912_EN_V1.0_07-2021_IFU_CE-MDR...
  • Page 19 Subsidiaries Instructions for use • Micromanipulator Interface VITOM 3D • MIC912_EN_V1.0_07-2021_IFU_CE-MDR...