Cleaning and Disinfecting

Warnings:
- Cleaning and disinfecting are separate processes. Cleaning must
precede disinfection. Before use by a different individual: clean,
disinfect, and check product for proper functioning.
- Clean the product regularly and when it becomes soiled.
- Follow safety guidelines on the chemical containers.
AGILITY Component
Back shell and hardware:
Cover:
Notes:
- Remove the foam pad from
the cover prior to cleaning and
disinfecting.
Secure the hook and loop fastener
strips at both ends of the cover.
- Turn the cover inside out.
- Zip the cover closed to avoid snags.
Foam pad, foam insert and lumbar
support pad:
Note: Remove the foam pad from the
cover prior to cleaning.
Care Symbols
Wipe with a damp cloth.
Hand wash.
Hand wash warm.
ROHO AGILITY Active Aluminum Back Series Operation Manual
To Clean:
Hand wash in warm (40°C) water with soap and a
sponge. Rinse with clean water. Wipe
dry with a clean cloth and then allow
to air dry completely.
Machine wash in warm (40°C) water with mild
detergent. Tumble dry low.
Wipe lightly with a damp cloth. DO NOT use soap. DO
NOT submerge in water. Wipe off with a clean cloth.
Completely air dry before replacing in cover.
Machine wash warm, normal.
40C
Dry flat in the shade.
Do not bleach.
Cautions:
- DO NOT use abrasives such as steel wool or scouring pads on the product.
- DO NOT use cleaning products containing petroleum or organic solvents [i.e.
acetone, toluene, Methyl Ethyl Ketone (MEK), naphtha, dry cleaning fluids,
adhesive removers] on the product.
To Disinfect:
Wipe gently with a cloth dampened with a household
disinfectant. Rinse with clean water. Wipe dry with a clean
cloth and then allow to air dry completely.
Hand wash using quaternary ammonium disinfectant (Follow
product label instructions). Rinse thoroughly with clean water.
The cover may also be machine washed in
40C
hot (60°C) water to disinfect. Tumble dry
low. DO NOT use bleach.
Note: The method of disinfection should be
determined based on disinfection needs.
Note: Disinfectants are NOT effective on porous surfaces
such as foam. Replace soiled or contaminated foam
components with authorized foam components only.
Replacement foam components may be purchased
separately. Soiled or contaminated foam components MUST
NOT be used by different individuals.
60C
15
60C
Machine wash hot, normal, at the temperature
shown.
Tumble dry low.