Service Menu Serv; Maintenance; Operation/Maintenance; Replacement Of The Ba�Ery - DEGA UKA III Instruction Manual

Table of Contents

Advertisement

The entry RELE is used for direct se�ng of the controller outputs. Entering this entry saves the current se�ing of the outputs,
which is later modified by the user. The controller outputs (RE1 to RE5, LIGH, SOUN, HV) can be browsed with LEFT and RIGHT
and pressing ENTER will switch over the output state. The outputs will remain in these se�ngs un�l we press ENTER on the
last EXIT entry. That will restore the output state, which existed before the user change.
The entry TRAN is similar to the entry RELE, except for the fact that the user sets the concentration on the given channel. Use
LEFT and RIGHT to select a channel and then press ENTER to enter the edit mode. Here we can use the arrows to change the
concentration ranging from minimal to maximal value of the configured transmi�er. The controller reacts to this set
concentration as if it was directly measured, which leads to the ac�va�on of alarms and switching of outputs.
The entry _RST resets the controller.

Service Menu SERV

The entry SERV ALL SENS makes the controller go into service mode. The communica�on with transmitters is terminated.
The outputs remain in the last known se�ng. Use LEFT and RIGHT buttons to set the service interval in a range of 1 to 24
hours. Press ENTER to start it's own regime.
The main segments display a countdown of the set interval in minutes, the time display shows the interval in hours, minutes
and seconds. After the countdown has elapsed, the controller returns to the detec�on mode. You can cancel this mode at any
�me by pressing the ENTER bu�on. Use LEFT and RIGHT keys to extend and shorten the interval after 10 minutes.
The entry SERV SEL SENS causes a blockage of any transmitter. Use LEFT and RIGHT to select a transmitter and ENTER to
block/unblock it. Blocked transmitters are indicated with an inscription DIS on the main segments.

Maintenance

1. Operation/Maintenance

In case of contamination the surface can be cleaned with a slightly moistened cloth. The connected transmitters require
performing of regular calibra�ons.
Recommended Default set�ngs:
1 x every 6 months carry out a „calibration" - adjust the sensitivity of the sensor using calibra�on gas and check the
func�onality of the system. The exact interval depends on the purity of the environment, required accuracy and the
occurrence of disturbing gases in the atmosphere.
The calibration interval can be changed by the DEGA Config software.
Perform calibration only at certified service centers with a valid cer�ficate of competence, or the manufacturer.
For the Czech Republic only DEGA CZ s.r.o.
2. Replacement of the ba�ery
The lifetime of the ba�ery in the controller is approximately 5 years. After
Energizer
2032
this time some func�ons of the controller may not work properly.
Duracell
2032
Remove the battery from the holder and replace it with one of the
Varta
2032
recommended types. Replacing the battery in the controller, which is not
Panasonic 2032
connected to the power supply, will erase the internal clock.
12