Deactivate peak hold
*Peak hold will not deactivate when the dynamometer is turned
Activate beep function
*Beep-function is only functional when connected to handheld display

6. FAQ

Anomalies
The device does not turn on
Display shows an incorrect
value
Display value is not reset to 0
Press and hold MENU until the display
shows 1.
Select 1 by pressing ENTER.
The display now shows
Press
Press ENTER to turn off Peak Hold.
Press RETURN to go to default settings.
Press and hold MENU until the display
shows 1.
Select 3 by pressing ENTER.
Use arrows to type in load limit in kg.
Press ENTER to save.
Press RETURN to go to default settings.
Possible causes
Discharged batteries
Electronics defective
More than 1 year since last
calibration
Malfunction of electronics
Has been overloaded
Has not been reset
- 30 -
PE
on.
until the display shows ∙PEoFF.
Corrective actions
Charge the battery
Contact Vetec's service depart-
ment
Re-calibrate at Vetec service
department
Reset: Turn off and on and retest
Contact Vetec's service depart-
ment