Time Out Events
Action
Auto exit error screen and power off
Auto skip error message screen and enter warmup
Auto exit menu and turn off Icon LED
Auto exit Force bump and Calibration
Exit auto detected span gas
Pairing, bump, and calibration result display
BLE pairing timeout
Troubleshooting
Problem
Battery icon blinks for 5
seconds when press
button to power on.
The detector, side LEDs, all
bays, and IntelliFlash blink
for 5 seconds when press
the button to power on.
All bays and IntelliFlash
light for 5 seconds
The detector, side LEDs,
and IntelliFlash light for 5
seconds, and sound two
long beeps.
The detector, side LEDs,
and IntelliFlash light for 5
seconds, and sound five
short beeps.
The detector, side LEDs,
and IntelliFlash light for 5
seconds, and sound one
long beep and two short
BW Icon
Shop for Gas products online at:
Cause
Depleted battery
The detector expired
All sensors fail
RTC fail
Data flash fail
Temperature sensor fail
CanarySense
www.
Charge the rechargeable
battery pack
The apparatus is over two
years lifetime, cannot
continue to use.
Replace the sensor or the
PCBA
Replace PCBA
Replace PCBA
Replace PCBA
34
.com
Time Out
5 seconds
5 seconds
6 seconds
30 seconds
60 seconds
5 seconds
60 seconds
Solution