Zero Calibration; Capture Real Time Reading; Set The Detector Via Device Configurator - Honeywell BW Icon User Manual

Portable multiple gas detector
Hide thumbs Also See for BW Icon:
Table of Contents

Advertisement

Zero Calibration

1. Double press the button to enter the menu.
2. Single press to switch to the calibration icon.
3. Press and hold the button for 3 seconds.
Zero calibration starts automatically, and the sensor LEDs light clockwise in blue.
After Zero calibration passes, the sensor LED lights solid blue for 5 seconds, and then the
detector is back to the normal mode.

Capture Real Time Reading

1. Pair your BW Icon with a mobile device.
2. In your mobile device, open the Device Configurator app.
3. Tap Menu
4. Tap Measurements
5. Tap Start Recording.

Set the Detector via Device Configurator

1. Pair the BW Icon with the Device Configurator App on your mobile device.
2. Tap the menu button
3. Tap Device Setup
4. Tap Download, to get the configuration table.
5.
Tap Edit to change the settings, and then tap Upload to apply them.