In terms of North America flammable gas performance approval:
Honeywell BW™ Icon and BW Icon+ are performance tested only for the range 0-5% methane in
the air as 0-100% LEL based on CSA 60079-29-1 and UL 60079-29-1
Only Honeywell BW™ Icon and BW Icon+ infrared sensor were evaluated for CSA 60079-29-1
and UL 60079-29-1.
The evaluation is valid with flow rate 500ml/min, CH4 (Methane) gas, and Safety Suite Device
Configurator in CSA lab test.
The other options are not the scope of CSA 60079-29-1.
For the compliance of CSA 60079-29-1, the adjustable alarm point shall not exceed 5% to 60 %.
The highest alarm shall be configured as a latching alarm, and the user can turn on/off the
latching alarm by Safety Suite Device Configurator or Device Configurator.
Honeywell BW ™ Icon and Icon+ were pressure tested for 80 to 120 kPa, temperature tested for -
40°C to 60°C, humidity tested for 5% to 90% RH, gas tested for 2.5%VOL CH4=50%LEL and air
velocity less than 6m/s in CSA lab test.
The battery voltage is 3.7V, and the manufacturer verifies the duration time until the low battery
condition mentioned; the performance test in the CSA lab only verifies the low battery duration.
The maximum power consumption of BW™ Icon and Icon+ are 380mW.
Infrared CH4 sensor warm-up time is less than 45s, CSA lab calibrates after warm-up for 1 hour,
and test gas application time is 60s. T90<60s for 50%LEL CH4 gas in diffusion mode.
To check reaction time, follow the section of 'Capture Real-Time Reading' to enter
'Measurement' in DC and apply calibration gas. The reaction time starts from the time once
attached hose or applied gas and end when reading over 90% of calibration gas concentration.
Performance Test Temperature Dependence:
-20 to 60°C, ±5%LEL or ±10% of reading at 20°C, whichever value is greater
-40 to -21°C, ±10%LEL or ±20% of reading at 20°C, whichever value is greater
Reading shows 0%LEL below 3%LEL and indicates Negative alarm once reading below -
5%LEL. Use the utility of manufacture to disable the suppression of reading.
Where it is necessary to apply LFL and UFL values for CSA 60079-29-1 and UL 60079-29-1,
reference shall be made to ANSI/NFPA 497.