Honeywell BW Icon User Manual page 16

Portable multiple gas detector
Hide thumbs Also See for BW Icon:
Table of Contents

Advertisement

3. Connect the instrument to the PC via the IR link.
4. Log in to SSDC with an authorized user account. For further information, refer to the SSDC
User Manual.
5. Click on the Device List View tab, the SSDC scans for connected devices or you can click
Refresh to browse manually.
6. Select the connected detector and then click Start Bump/Cal.
7. In the Start Bump/Calibration Test window, do the following:
Select Calibration;
l
Select the calibration sensor. You can modify the default values;
l
Click START TEST
l
Wait for several seconds. The detector starts ZERO  Calibration, and the four slots
l
LEDs flash blue clockwise. After Zero Calibration is complete, the LEDs are solid blue
if calibration passed, or red if failed.
8. Optional Step. You can either remove or keep the IR link connection, the remaining
operation is in the instrument.
9. Apply span gas when the slot LEDs start flashing. Span calibration starts after the instrument
detects gas. The four slots LEDs flash blue clockwise. After the Span calibration is completed,
the LEDs are solid green if calibration passed, or red if failed.
10. Remove the calibration cap. The detector starts purging, and the slots LEDs flash in yellow
clockwise.
After the purge is complete, the device is back to the regular mode.
BW Icon
Shop for Gas products online at:
CanarySense
www.
18
.com
User Manual
1.888.610.7664