Mirion Technologies TelePole II User Manual

Mirion Technologies TelePole II User Manual

Telescopic meter

Advertisement

 
 
TelePole II Telescopic Meter
Prepared by:
Name
Date
Signature
All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,
reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
 
 
User's Guide
Reviewed by:
15‐00143 
Revision 1  
April 2016
Approved by:
 
 

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TelePole II and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Mirion Technologies TelePole II

  • Page 1 Approved by: Name Date Signature All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 2         All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 3 RMVC software        All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 4: Table Of Contents

            All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 5 6.7  Firmware Upgrade       All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 6: The Telepole Ii Meter

        Figure 1‐1. TelePole II Meter  All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 7: Features

     Also supports ALARA in that the operators can see their local dose as well as the remote  dose from the external detector.  All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 8: Telepole General Specifications

    85 hours of continuous operation     Temperature range Operation:   ‐20C to +50C (‐4F to 122F)  All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 9         All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 10: Telepole Detector Specifications

    Weight:   220 grams (0.5 lbs)      All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 11 The increased change in the response time depends on the current reading and  the extent of the change in the radiation field. Generally it will be less than 2 sec.    All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 12: Operating Instructions

    Loctite  221  (or  similar)  compound  for  when  the  components are reattached.  All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 13   Figure 4‐3. Meter Connection to Pole     All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 14: Starting-Up

            All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 15: The Display

            All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 16 LED (LED Flashlight) Mode  This command activates the LED on the tip of the detector.     Push MENU to see the next Line…   All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 17 Section 6.6 for more details.  OR  All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 18: General Functions

    Internal Detector Range: 0.01 mR/h to 1.5 R/h.    Readings are displayed digitally and updated every second    All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 19         All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 20: Threshold Selection

    4.5.4  Both these thresholds are set using the external RMV software which available  off our website.     All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 21: Push-Buttons Function

            All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 22: Meter Display

          All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 23         All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 24: Communication

    Currently, the Bluetooth communication is only setup to be activated during an Alarm  condition.  The first step is to pair the earphone (earphones that require a PIN for activation  will not connect to the Telepole II) or the Mirion Vibrating Bluetooth Wristband to the  TelePole II. This is achieved by activating the pairing procedure on the earphone/bracelet  All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 25         All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 26   6.4.5. NOTE – if this is the first time connecting the TelePole II to the user computer, you  will need to establish which com port is to be used.  To do this go to Device  All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 27 Click Connect.  6.4.8. Prior to clicking on RMC, ensure that the TelePole II is communicating to the  software (the dose and rate information is displayed along with the serial number  All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 28 Due will track when the Calibration is Due and show a Calibrate Alarm (D‐CAL)  when overdue.    All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 29 Click the save icon and name for later use     All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 30 ** The tight range for the F3 value is to provide good linearity for the High Range tube in  the switch – over range 1500 – 400 mR/h.  Higher calibration points may be used if the  customer verifies acceptability of high range response in switchover range.  All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 31 6.5.12. To adjust the “F” Factors – click the calibrate button.    All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 32 Calibration Factors (F Values) are completed, complete the calibration by updating  the remaining Values in the Cal. Point Column.  (Expose the meter / detector to the  appropriate value and click in the Cal. Point Block of that point to update.)    All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 33 6.6.6  Repeat the previous step for the remaining F Values and then complete an additional  linearity point needed for the As Left dose rates.      All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 34 6.7.6  Then select Send File.  It may take a minute or so to begin loading (a status bar at the  bottom of the window will show a green bar moving across during the loading process).   Once successful you should see a popup window:    All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.
  • Page 35   6.7.9  Repeat the process with the External Detector if needed.     All of the information in this document is the exclusive property of Mirion Technologies (MGPI), Inc. and is not to be disclosed,   reproduced, or used except as authorized in writing by Mirion Technologies (MGPI), Inc.

Table of Contents