Page 2
The manufacturer reserves the right to modify technical data without prior notice. The Endress+Hauser sales organization will supply you with current information and updates to these instructions. Endress+Hauser...
Micropilot FMR62B PROFINET with Ethernet-APL Table of contents Table of contents 5.3.4 Antenna, PTFE cladded, flush mount About this document ....5 80 mm (3 in) ....
Page 4
Table of contents Micropilot FMR62B PROFINET with Ethernet-APL 11.7 Event logbook ......58 System integration ....41 11.7.1 Event history .
Micropilot FMR62B PROFINET with Ethernet-APL About this document About this document Purpose of this document These Operating Instructions contain all the information that is required in the various phases of the life cycle of the device: from product identification, incoming acceptance and storage, to mounting, connection, operation and commissioning through to troubleshooting, maintenance and disposal.
Registered trademark of the PROFIBUS User Organization, Karlsruhe, Germany Bluetooth® The Bluetooth® word mark and logos are registered trademarks owned by the Bluetooth SIG, Inc. and any use of such marks by Endress+Hauser is under license. Other trademarks and trade names are those of their respective owners. Apple®...
Page 7
Micropilot FMR62B PROFINET with Ethernet-APL About this document KALREZ®, VITON® Registered trademarks of DuPont Performance Elastomers L.L.C., Wilmington, DE USA Endress+Hauser...
Do not touch or clean device surfaces with pointed or hard objects. Clarification of borderline cases: ‣ For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in verifying the corrosion resistance of fluid-wetted materials, but does not accept any warranty or liability.
It meets the general safety standards and legal requirements. It also complies with the EU directives listed in the device-specific EU Declaration of Conformity. Endress+Hauser confirms this by affixing the CE mark to the device. IT security We only provide a warranty if the device is installed and used as described in the Operating Instructions.
Basic safety instructions Micropilot FMR62B PROFINET with Ethernet-APL • Write protection via hardware write protection switch • Access code to change user role (applies to operation via display, Bluetooth or FieldCare, DeviceCare, asset management tools (e.g., AMS, PDM and Web server)
Page 11
Micropilot FMR62B PROFINET with Ethernet-APL Basic safety instructions Description of device parameters. Endress+Hauser...
All the information about the measuring device is displayed. ‣ Endress+Hauser Operations app; manually enter the serial number indicated on the nameplate or scan the 2D matrix code on the nameplate. All the information about the measuring device is displayed.
Incoming acceptance and product identification Micropilot FMR62B PROFINET with Ethernet-APL 4.2.2 Manufacturer address Endress+Hauser SE+Co. KG Hauptstraße 1 79689 Maulburg, Germany Place of manufacture: See nameplate. Storage and transport 4.3.1 Storage conditions • Use the original packaging • Store the device in clean and dry conditions and protect from damage caused by shocks Storage temperature range See Technical Information.
Micropilot FMR62B PROFINET with Ethernet-APL Mounting Mounting General instructions WARNING Loss of protection rating if the device is opened in a wet environment. ‣ Only open the device in a dry environment! 1. Install the device or turn the housing so that the cable entries do not point upwards.
Mounting Micropilot FMR62B PROFINET with Ethernet-APL 5.2.2 Avoiding interference echoes A0031813 Metal deflector plates, installed at an angle to scatter the radar signals, help prevent interference echoes. 5.2.3 Vertical alignment of antenna axis Align the antenna so that it is perpendicular to the product surface.
Micropilot FMR62B PROFINET with Ethernet-APL Mounting Maximum nozzle length H as a function of the nozzle diameter D D 80 to 100 mm (3.2 to 4 in) 1 700 mm (67 in) 100 to 150 mm (4 to 6 in) 2 100 mm (83 in) ≥...
Mounting Micropilot FMR62B PROFINET with Ethernet-APL 5.3.3 Antenna, PTFE cladded, flush mount 50 mm (2 in) The cladding on the antenna also acts as a process seal. An additional seal is not required for installation. Mounting cladded flanges Note the following for cladded flanges: •...
Page 19
Micropilot FMR62B PROFINET with Ethernet-APL Mounting Mounting cladded flanges Note the following for cladded flanges: • Use the same number of flange screws as the number of flange bores provided. • Tighten the screws with the necessary torque (see Table).
Mounting Micropilot FMR62B PROFINET with Ethernet-APL The maximum length of the nozzle H depends on the nozzle diameter D D 80 to 100 mm (3.2 to 4 in) 1 750 mm (70 in) 100 to 150 mm (4 to 6 in) 2 200 mm (88 in) ≥...
Micropilot FMR62B PROFINET with Ethernet-APL Mounting 4 0.7 Nm A0038224 1. If fitted: release the screw of the cover lock for the electronics compartment cover using the Allen key. 2. Unscrew the electronics compartment cover from the transmitter housing and check the cover seal.
Page 22
Mounting Micropilot FMR62B PROFINET with Ethernet-APL WARNING Supply voltage switched on! Risk of electric shock and/or explosion! ‣ Switch off the supply voltage before opening the measuring device. 1 → 2 → 3 → 4 mm 0.7 Nm A0046831 A0046832 A0046833 ‣...
Micropilot FMR62B PROFINET with Ethernet-APL Mounting 7 → A0048406 A0046928 ‣ ‣ Plug in the connection for the Screw the display cover firmly display module in the back onto the housing. If fitted: connection compartment. tighten the screw of the cover ‣...
Electrical connection Micropilot FMR62B PROFINET with Ethernet-APL Electrical connection Connecting requirements 6.1.1 Cover with securing screw In the case of devices for use in the hazardous area with a certain type of protection, the cover is sealed by a securing screw.
Electrical connection Micropilot FMR62B PROFINET with Ethernet-APL 6.2.2 Cable specification Rated cross-section • Supply voltage 0.5 to 2.5 mm (20 to 13 AWG) • Protective earth or grounding of the cable shield > 1 mm (17 AWG) • External ground terminal 0.5 to 4 mm...
Micropilot FMR62B PROFINET with Ethernet-APL Electrical connection Overvoltage category Overvoltage category II 6.2.4 Wiring WARNING Supply voltage might be connected! Risk of electric shock and/or explosion! ‣ If the device is used in hazardous areas, make sure to comply with national standards and the specifications in the Safety Instructions (XAs).
Electrical connection Micropilot FMR62B PROFINET with Ethernet-APL Devices with M12 plug A0011175 7 View of the plug-in connection on the device APL signal - APL signal + Shielding Not assigned Various M12 sockets are available as accessories for devices with M12 plugs.
• Operation via Bluetooth® wireless technology (with optional device display with Bluetooth) with SmartBlue app or FieldXpert, DeviceCare • Operation via Web server • Operation via operating tool (Endress+Hauser FieldCare/DeviceCare) or FDI Hosts (e.g., PDM) Operating keys and DIP switches on the electronic...
• Press and hold the keys simultaneously to return to the upper level 7.4.2 Operation via Bluetooth® wireless technology (optional) Prerequisite • Device with device display including Bluetooth • Smartphone or tablet with Endress+Hauser SmartBlue app or PC with DeviceCare from version 1.07.05 or FieldXpert SMT70 Endress+Hauser...
Micropilot FMR62B PROFINET with Ethernet-APL Operation options The connection has a range of up to 25 m (82 ft). The range can vary depending on environmental conditions such as attachments, walls or ceilings. The operating keys on the display are locked as soon as the device is connected via Bluetooth.
Operation options Micropilot FMR62B PROFINET with Ethernet-APL 7.5.2 Requirements Computer software Recommended operating systems • Microsoft Windows 7 or higher. • Mobile operating systems: • iOS • Android Microsoft Windows XP is supported. Web browsers supported • Microsoft Internet Explorer 8 or higher •...
Micropilot FMR62B PROFINET with Ethernet-APL Operation options 7.5.3 Establishing a connection Via PROFINET with Ethernet-APL network A0046097 10 Options for remote operation via PROFINET with Ethernet-APL network: star topology Automation system, e.g., Simatic S7 (Siemens) Ethernet switch Computer with Web browser (e.g., Microsoft Edge) for accessing the integrated device Web server or computer with operating tool (e.g., FieldCare, DeviceCare, SIMATIC PDM) with iDTM Profinet Communication...
Operation options Micropilot FMR62B PROFINET with Ethernet-APL A0046626 11 Web browser login Device tag Device name Status signal Locking status Current measured values Select the language Enter the "Password" parameter Login 1. Select the preferred Language parameter for the Web browser.
Micropilot FMR62B PROFINET with Ethernet-APL Operation options Header The following information appears in the header: • Device tag • Device name • Status signal • Locking status • Current measured values Navigation area If a function is selected in the function bar, the submenus of the function open in the navigation area.
Operation options Micropilot FMR62B PROFINET with Ethernet-APL Range of functions of "Web server functionality" parameter Option Description Disable • The Web server is completely disabled. • Port 80 is locked. Enable • The complete functionality of the Web server is available.
Micropilot FMR62B PROFINET with Ethernet-APL Operation options 7.6.1 Connecting the operating tool Via PROFINET with Ethernet-APL network A0046097 14 Options for remote operation via PROFINET with Ethernet-APL network: star topology Automation system, e.g., Simatic S7 (Siemens) Ethernet switch Computer with Web browser (e.g., Microsoft Edge) for accessing the integrated device Web server or computer with operating tool (e.g., FieldCare, DeviceCare, SIMATIC PDM) with iDTM Profinet Communication...
7.8.1 Function scope Tool for connecting and configuring Endress+Hauser field devices. The fastest way to configure Endress+Hauser field devices is with the dedicated "DeviceCare" tool. Together with the device type managers (DTMs), DeviceCare presents a convenient, comprehensive solution. For details, see Innovation Brochure IN01047S...
Micropilot FMR62B PROFINET with Ethernet-APL System integration System integration Overview of device description files 8.1.1 Current version data for the device Firmware version 01.00.zz • On the title page of the Operating Instructions • On the transmitter nameplate • System → Information → Firmware version Firmware version parameter release date 04.2023...
System integration Micropilot FMR62B PROFINET with Ethernet-APL 8.2.1 File name of the device master file (GSD) Example of the name of a device master file: GSDML-V2.43-EH-Micropilot-202304dd.xml (dd corresponds to the release day) GSDML Description language V2.43 Version of the PROFINET specification...
Micropilot FMR62B PROFINET with Ethernet-APL System integration Cyclic data transmission 8.3.1 Overview of the modules The following graphic shows which modules are available to the device for cyclic data exchange. Cyclic data exchange is performed with an automation system. Application → PROFINET → Analog input → Analog input 1 to 11 → Assign process...
System integration Micropilot FMR62B PROFINET with Ethernet-APL Module: Binary input The Binary input module can cyclically send discrete values from the device to the automation system. The status of the Heartbeat Verification is transmitted for the device: Module: Binary input Heartbeat Technology slot 80...
Micropilot FMR62B PROFINET with Ethernet-APL System integration 8.3.4 Startup configuration Startup configuration The automation system adopts the configuration of the most important parameters of the device. (NSU) • Interfaces: • Display operation • Web server functionality • Bluetooth activation • Service (UART-CDI) •...
Commissioning Micropilot FMR62B PROFINET with Ethernet-APL Commissioning All configuration tools provide a commissioning assistant that supports the user when setting the most important configuration parameters (Guidance menu Commissioning wizard). Preliminaries The measuring range and the unit in which the measured value is transmitted correspond to the data on the nameplate.
Via service interface (CDI) A0039148 Computer with FieldCare/DeviceCare operating tool Commubox FXA291 Service interface (CDI) of the measuring device (= Endress+Hauser Common Data Interface) 9.3.3 Hardware settings Setting the device name A measuring point can be quickly identified within a plant on the basis of the Device tag parameter and PROFINET device name parameter.
Commissioning Micropilot FMR62B PROFINET with Ethernet-APL Configuring the operating language 9.4.1 Local display Setting the operating language To set the operating language, the display first needs to be unlocked: 1. Press the key for at least 2 s. A dialog box appears.
Micropilot FMR62B PROFINET with Ethernet-APL Commissioning Configuring the device 9.5.1 Level measurement in liquids 100% A0016933 17 Configuration parameters for level measurement in liquids Reference point of measurement Length of antenna + 10 mm (0.4 in) 50 to 80 mm (1.97 to 3.15 in); medium εr < 2...
Commissioning Micropilot FMR62B PROFINET with Ethernet-APL Recording the Echo curve Record the current Echo curve as a reference echo curve for subsequent diagnostic purposes After the measurement has been configured, it is recommended to record the current Echo curve as a reference echo curve.
Micropilot FMR62B PROFINET with Ethernet-APL Operation Operation 10.1 Reading off the device locking status Displaying active write protection: • In the Locking status parameter Menu path of local display: at the top operating level Menu path of operating tool: System → Device management •...
Diagnostics and troubleshooting Micropilot FMR62B PROFINET with Ethernet-APL Diagnostics and troubleshooting 11.1 General troubleshooting 11.1.1 General errors Device is not responding • Possible cause: Supply voltage does not match the specification on the nameplate Remedial action: Apply the correct voltage •...
Remedial action: The standard user name is "admin" and the password is the device serial number indicated on the device nameplate (only if the password was not changed by the user beforehand) If the password has been forgotten, contact Endress+Hauser Service (www.addresses.endress.com) Connection via SmartBlue not possible •...
Diagnostics and troubleshooting Micropilot FMR62B PROFINET with Ethernet-APL 11.3.1 Diagnostic information on local display A0043759 19 Segment display without keys Status symbol for event level Status signal with diagnostic event X X X X X X X S801 X X X X X X X A0043103 ...
Micropilot FMR62B PROFINET with Ethernet-APL Diagnostics and troubleshooting 11.3.4 Diagnostic event and event text The fault can be identified by means of the diagnostic event. The event text helps you by providing information about the fault. In addition, the associated status symbol is displayed in front of the diagnostic event.
Diagnostics and troubleshooting Micropilot FMR62B PROFINET with Ethernet-APL 2 → Diagnostic list Diagnostics 1 S801 Supply voltage Diagnostics 2 Diagnostics 3 A0051132 ‣ Select the diagnostic event and confirm Supply voltage (ID:203) S801 0d00h02m25s Increase supply voltage A0051133 ‣ Remedy informationClosing 11.4.2...
Page 57
Micropilot FMR62B PROFINET with Ethernet-APL Diagnostics and troubleshooting Diagnostic Short text Remedy instructions Status Diagnostic number signal behavior [from the [from the factory] factory] Build-up detected Check process conditions Warning Diagnostic of electronic Real time clock Replace main electronics Warning...
Diagnostics and troubleshooting Micropilot FMR62B PROFINET with Ethernet-APL Diagnostic Short text Remedy instructions Status Diagnostic number signal behavior [from the [from the factory] factory] Record map Recording of mapping Warning please wait Diagnostic of process Supply voltage too low Increase supply voltage...
Micropilot FMR62B PROFINET with Ethernet-APL Diagnostics and troubleshooting 11.7.2 Filtering the event logbook Filters can be used to determine which category of event messages is displayed in the Event list submenu. Navigation: Diagnostics → Event logbook Filter categories • All •...
Diagnostics and troubleshooting Micropilot FMR62B PROFINET with Ethernet-APL Info number Info name I1666 Clock synchronized I1712 New flash file received I1956 Reset 11.8 Resetting the device 11.8.1 Reset password via operating tool Enter a code to reset the current ' M aintenance' password.
Micropilot FMR62B PROFINET with Ethernet-APL Diagnostics and troubleshooting Resetting the device to the factory setting 12 s Display 50 ms A0050015 22 Sequence - Factory setting Reset the device to the factory setting ‣ Press operating key I for at least 12 s.
Maintenance Micropilot FMR62B PROFINET with Ethernet-APL Maintenance No special maintenance work is required. 12.1 Exterior cleaning Notes on cleaning • The cleaning agents used should not corrode the surfaces and the seals • Observe the degree of protection of the device 12.2...
Repair concept Endress+Hauser repair concept • The devices have a modular design • Repairs can be carried out by Endress+Hauser Service or by appropriately trained customers • Spare parts are grouped into logical kits with the associated replacement instructions For more information on service and spare parts, please contact your Endress+Hauser sales representative.
Repair Micropilot FMR62B PROFINET with Ethernet-APL 13.3.1 HistoROM It is not necessary to perform a new device calibration after replacing the display or transmitter electronics. The spare part is supplied without HistoROM. After removing the transmitter electronics, remove HistoRom and insert it into the new spare part.
Micropilot FMR62B PROFINET with Ethernet-APL Accessories Accessories 14.1 Weather protection cover 316L The weather protection cover can be ordered together with the device via the "Accessory enclosed" product structure. It is used to protect against direct sunlight, precipitation and ice.
Accessories Micropilot FMR62B PROFINET with Ethernet-APL Connecting cable: • Connecting cable (option) up to 30 m (98 ft) • Customer-supplied standard cable up to 60 m (197 ft) Recommendation: EtherLine®-P CAT.5e from LAPP. Specification of customer-supplied connecting cable Push-in CAGE CLAMP®, connection technology, push actuation •...
Micropilot FMR62B PROFINET with Ethernet-APL Technical data Technical data 15.1 Input Measured variable The measured variable is the distance between the reference point and the product surface. The level is calculated based on "E", the empty distance entered. Measuring range The measuring range starts at the point where the beam hits the tank floor.
Page 70
Technical data Micropilot FMR62B PROFINET with Ethernet-APL 100% A0051658 28 Usable measuring range Length of antenna + 10 mm (0.4 in) Usable measuring range 50 to 80 mm (1.97 to 3.15 in); medium εr < 2 Vessel height Reference point of measurement, varies depending on the antenna system For further information on the reference point, see →...
Page 71
For the dielectric constants (DC values) of many media commonly used in industry, please refer to: • Dielectric constant (DC value) Compendium CP01076F • The Endress+Hauser "DC Values app" (available for Android and iOS) Measurement in storage vessel Storage vessel - measuring conditions Calm medium surface (e.g.
Page 72
Technical data Micropilot FMR62B PROFINET with Ethernet-APL Antenna, PTFE cladded flush mount, 50 mm (2 in) in storage vessel Media group Measuring range A0 (ε 1.2 to 1.4) 7 m (23 ft) A (ε 1.4 to 1.9) 12 m (39 ft) B (ε...
Page 73
Micropilot FMR62B PROFINET with Ethernet-APL Technical data PTFE Drip-off antenna, 50 mm (2 in) in buffer vessel Media group Measuring range A0 (ε 1.2 to 1.4) 4 m (13 ft) A (ε 1.4 to 1.9) 7 m (23 ft) B (ε...
Page 74
Technical data Micropilot FMR62B PROFINET with Ethernet-APL Antenna, horn 316L, 65 mm (2.6 in) in buffer vessel Media group Measuring range A0 (ε 1.2 to 1.4) 11 m (36 ft) A (ε 1.4 to 1.9) 21 m (69 ft) B (ε...
Page 75
Micropilot FMR62B PROFINET with Ethernet-APL Technical data Antenna, PTFE cladded flush mount, 80 mm (3 in) in vessel with agitator Media group Measuring range A0 (ε 1.2 to 1.4) 7 m (23 ft) A (ε 1.4 to 1.9) 13 m (43 ft) B (ε...
Technical data Micropilot FMR62B PROFINET with Ethernet-APL Measurement in bypass Bypass measuring conditions Application in vessels with moving medium surface (e.g. permanent free filling from above, mixing jets) Antenna, PTFE cladded flush mount, 80 mm (3 in) in bypass Media group Measuring range A0 (ε...
Page 77
Micropilot FMR62B PROFINET with Ethernet-APL Technical data Pre-programmed linearization curves Linearization tables for calculating the volume in the following vessels are preprogrammed into the device: • Pyramid bottom • Conical bottom • Angled bottom • Horizontal cylinder • Sphere Other linearization tables of up to 32 value pairs can be entered manually.
Technical data Micropilot FMR62B PROFINET with Ethernet-APL Supported functions • Identification & maintenance Simple device identification via: • Control system • Nameplate • Measured value status The process variables are communicated with a measured value status • Blinking feature via the local display for simple device identification and assignment •...
Page 79
Micropilot FMR62B PROFINET with Ethernet-APL Technical data Plastic housing Plastic housing; process temperature –20 to +150 °C (–4 to +302 °F) A0032024 29 Plastic housing; process temperature –20 to +150 °C (–4 to +302 °F) = T : –20 °C (–4 °F) | T : +76 °C (+169 °F)
Page 80
Technical data Micropilot FMR62B PROFINET with Ethernet-APL Plastic housing; process temperature –20 to +200 °C (–4 to +392 °F) A0032024 31 Plastic housing; process temperature –20 to +200 °C (–4 to +392 °F) = T : –20 °C (–4 °F) | T : +76 °C (+169 °F)
Page 81
Micropilot FMR62B PROFINET with Ethernet-APL Technical data Restriction to a process temperature of 0 to +150 °C (+32 to +302 °F) with CSA C/US approval and plastic housing A0048826 34 Plastic housing; process temperature 0 to +150 °C (+32 to +302 °F) with CSA C/US approval = T...
Page 82
Technical data Micropilot FMR62B PROFINET with Ethernet-APL Plastic housing; process temperature –40 to +280 °C (–40 to +536 °F) A0032024 37 Plastic housing; process temperature –40 to +280 °C (–40 to +536 °F) = T : –40 °C (–40 °F) | T : +76 °C (+169 °F)
Page 83
Micropilot FMR62B PROFINET with Ethernet-APL Technical data Restriction to a process temperature of 0 to +450 °C (+32 to +842 °F) with CSA C/US approval and plastic housing A0048826 40 Plastic housing; process temperature 0 to +450 °C (+32 to +842 °F) with CSA C/US approval = T...
Page 84
Technical data Micropilot FMR62B PROFINET with Ethernet-APL Plastic housing; process temperature –196 to +200 °C (–320 to +392 °F) A0050248 43 Plastic housing; process temperature –196 to +200 °C (–320 to +392 °F) = T : –196 °C (–320 °F) | T : +76 °C (+169 °F)
Page 85
Micropilot FMR62B PROFINET with Ethernet-APL Technical data Aluminum housing; process temperature –20 to +200 °C (–4 to +392 °F) A0032024 46 Aluminum housing, coated; process temperature –20 to +200 °C (–4 to +392 °F) = T : –20 °C (–4 °F) | T : +79 °C (+174 °F)
Page 86
Technical data Micropilot FMR62B PROFINET with Ethernet-APL Aluminum housing; process temperature –40 to +280 °C (–40 to +536 °F) A0032024 49 Aluminum housing, coated; process temperature –40 to +280 °C (–40 to +536 °F) = T : –40 °C (–40 °F) | T : +79 °C (+174 °F)
Page 87
Micropilot FMR62B PROFINET with Ethernet-APL Technical data Aluminum housing; process temperature –196 to +200 °C (–320 to +392 °F) A0050248 52 Aluminum housing, coated; process temperature –196 to +200 °C (–320 to +392 °F) = T : –196 °C (–320 °F) | T : +79 °C (+174 °F)
Page 88
Technical data Micropilot FMR62B PROFINET with Ethernet-APL 316L housing; process temperature –40 to +150 °C (–40 to +302 °F) A0032024 55 316L housing; process temperature range: –40 to +150 °C (–40 to +302 °F) = T : –40 °C (–40 °F) | T : +77 °C (+171 °F)
Page 89
Micropilot FMR62B PROFINET with Ethernet-APL Technical data 316L housing; process temperature –40 to +450 °C (–40 to +842 °F) A0032024 58 316L housing; process temperature –40 to +450 °C (–40 to +842 °F) = T : –40 °C (–40 °F) | T : +77 °C (+171 °F)
Page 90
Technical data Micropilot FMR62B PROFINET with Ethernet-APL 316L housing, hygiene 316L housing, hygiene; process temperature –20 to +150 °C (–4 to +302 °F) A0032024 61 316L housing, hygiene; process temperature –20 to +150 °C (–4 to +302 °F) = T : –20 °C (–4 °F) | ...
Page 91
Micropilot FMR62B PROFINET with Ethernet-APL Technical data 316L housing, hygiene; process temperature –40 to +200 °C (–40 to +392 °F) A0032024 64 316L housing, hygiene; process temperature –40 to +200 °C (–40 to +392 °F) = T : –40 °C (–40 °F) | T : +76 °C (+169 °F)
Page 92
Technical data Micropilot FMR62B PROFINET with Ethernet-APL Climate class DIN EN 60068-2-38 (test Z/AD) Installation height as per Generally up to 5 000 m (16 404 ft) above sea level IEC61010-1 Ed.3 Degree of protection Test as per IEC 60529 and NEMA 250-2014...
Micropilot FMR62B PROFINET with Ethernet-APL Technical data 15.4 Process Process pressure range WARNING The maximum pressure for the device depends on the lowest-rated component with regard to pressure (components are: process connection, optional mounted parts or accessories). ‣ Only operate the device within the specified limits for the components! ‣...
Page 94
–20 to +200 °C (–4 to +392 °F) The pressure range may be further restricted in the event of a CRN approval. Dielectric constant For liquids ε ≥ 1.2 Contact Endress+Hauser for applications with lower dielectric constants than indicated. Endress+Hauser...
Need help?
Do you have a question about the Micropilot FMR62B PROFINET with Ethernet-APL and is the answer not in the manual?