After entering the sterilization confi rming interface, you must follow the prescribed steps to light
the start icon.
a) Open the inner door of the incubator for 30s until you hear the alarm sound;
b)Remove all items in the incubator (including but not limited to samples, HEPA fi lters, etc.);
c) Drain the water from the reservoir and wipe up any remaining water;
d)Close all doors of the incubator and click "Start".
Caution
Prior to sterilization, make sure that the samples and HEPA filter inside the
incubator have been taken out, otherwise may cause high temperature damage.
Warning Prior to sterilization, make sure to drain the water in the reservoir and wipe up any
remaining water; otherwise, the water inside the incubator will evaporate under high temperature
and then generate pressure, while high temperature moisture is harmful to human body;
Caution
The sterilization procedure cannot operate in the event of any of the following
faults:
a) Sensor damage
b) High actual value
c) Low actual value
d) Unreasonable actual value
e) Wrong communication
f) Water detected
Warning Hot surface, caution!
During high temperature sterilization, the glass door handle, glass door and lining surface of the
outer door will become very hot. Do not open the doors before the internal temperature cools
down to normal temperature. Wear safety gloves when contacting these surfaces during routine
sterilization.
During the sterilization process, opening the inner glass door will pause the sterilization. An
alarm will prompt the user to decide how to continue.