Disconnect tubes
Prime pump tubing
Process blank glass slides
®
Hematek
3000 System – Instruction Manual – Version 1.3
2. Reach underneath the platen through the waste tank area, see
Figure 8-5.
1
2
8
1 Spouts
2 Rinse tube
3 Buffer tube
4 Stain tube
5 Service button
6 Platen
7 Waste tank area
8 Slide drawer
Figure 8-5 Waste tank area, bottom view
3. Disconnect rinse, buffer, and stain tubes from the spouts under the
platen.
4. Press the service button.
5. Place the free tube ends into the beaker.
6. Prime the pump tubing, see Chapter 5.2.
7. Place the free ends of the stain and buffer tubes into separate 10mL
graduated cylinders.
8 Troubleshooting