Self Calibration

Calibrates the integrated front sensor based on the USB color sensor. Or to manually adjust the integrated front sensor with an
external near range color analyzer reference, which is traceable to a primary standards laboratory. All PICTURE MODES are
updated when Self Calibration is performed.
NOTE:
AUTO CALIB. mode requires connecting the USB color sensor to the sensor port.
EXPERT CALIB. mode requires an external near range color analyzer.
1. To show the CALIBRATION menu, plug-in the USB color sensor to the sensor port (Figure S.1) or select STAND ALONE
CALIBRATION in the advanced menu (page 22).
2. Use "LEFT" or "RIGHT" to select SELF in the MODE selection (Figure S.2). Press "UP" or "DOWN" to move to the next
selection.
Figure S.1
3. Use "LEFT" or "RIGHT" to select AUTO CALIB. or EXPERT CALIB. in the OPTION selection. Press SELECT to go to the
next section.
To select AUTO CALIB., go to step 4. To select EXPERT CALIB., go to step 5.
[AUTO CALIB.] Make sure that the USB color sensor is connected to the sensor port.
4. The procedure will ask for the USB color sensor to be placed on the center of the display panel (Figure S.3). Tilt the display
panel approximately 5° backward and place the USB color sensor in the center of the display panel (Figure S.1).
NOTE:
Place the USB color sensor fl at against the LCD to avoid any infl uence of ambient light.
DO NOT press the USB color sensor against the display panel.
NOTE:
Make sure that the USB color sensor cable is positioned to avoid blocking the ambient light and human sensors.
Press "SELECT" to start the calibration.
Go to step 8.
[EXPERT CALIB.] An external near range color analyzer, which is traceable to a primary standards laboratory, is required for this
method.
5. The procedure will ask for the external near range color analyzer to be placed on the center of the display panel. Place the
external near range color analyzer in the center of the display panel, then press SELECT.
6. The measument screen will be displayed. Measure the luminance and chromaticity value of the center of the display by
using the external near range color analyzer. Input the measured value to LUMINANCE and chromaticity by using the control
keys, then press SELECT (Figure S.4).
7. According to the message window, repeat the measurement of luminance and chromaticity and input the measured value
until the procedure is completed.
NOTE:
If chromaticity value is not measured by your external near range color analyzer, only input the luminance value and
press SELECT.
8. Gamma Adjust and DICOM Measurement are then executed automatically. After completing DICOM Measurement, you can
set DATE and TIME by using control keys.