 Temperature condition: 23℃±5℃
 Humidity condition: ≤65% R.H.
 Zero value adjustment : open circuit and short circuit zeroed before the test
 Warm-up time:> 30 minutes
 Calibration time: 12 months
C: 0.20%*(1+ Cx/Cmax+ Cmin/Cx )(1+Dx)(1+ks+kv+kf);
L: 0.20%*(1+ Lx/Lmax+ Lmin/Lx )(1+1/Qx)(1+ks+kv+kf);
Z: 0.20%*(1+ Zx/Zmax+ Zmin/Zx )(1+ks+kv+kf);
R: 0.20%*(1+ Rx/Rmax+ Rmin/Rx )(1+Qx)(1+ks+kv+kf);
D: 0.20%*(1+ Zx/Zmax+ Zmin/Zx )(1+Dx+Dx*Dx)(1+ks+kv+kf );
Q: 0.20%*(1+ Zx/Zmax+ Zmin/Zx )(Qx+1/Qx )(1+ks+kv+kf );
Wherein
1. L, C, R, Z are relative errors; D, Q, θ are absolute errors
2. Where the subscript is x, it is the measurement of this parameter; where the subscript is max, it
is the maximum value and where the subscript is min, it is the minimum value
3. Ks is the velocity factor, kv is the voltage factor, kf is the frequency factor
The maximum and minimum values of the measurement parameters that affect accuracy
Frequenc
100
120
y(Hz)
Cmax
800
667
Cmin
1500
1250
Lmax
1590
1325
Lmin
3.2
2.6
Z/Rmax
1
Z/Rmin
1.59
Frequenc
15K
y (Hz)
Cmax
5.3
Cmin
10
Lmax
10.6
Lmin
0.021
Z/Rmax
1
Z/Rmin
1.59
Wherein: the unit of Cmax is uF; the unit of Cmin is pF; the unit of Lmax is H; the unit of Lmin is
mH; the unit of Zmax / Rmax is MΩ; the unit of Zmin / Rmin is in Ω;
Measurement speed error factor ks:
Speed Mode
Slow
ks
0
Frequency
100~1K
kf
0
Measurement speed error factor kv:
Level(mV)
100
300
kv
10
3
200
400
800
400
200
100
750
375
187
795
397.5
198.8
1.6
0.8
0.4
20K
40K
4
2
7.5
3.75
7.95
3.98
0.016
0.008
Medium
Fast
1
8
2K~10K
15K
0.5
1
600
1000
1
0
1K
2K
80
40
150
75
159
79.5
0.32
0.16
50K
80K
1.6
1
3
1.87
3.18
1.99
0.0064
0.004
Measurement speed error factor kf:
20K
40K
50K
2
1500
2000
1
2
18
4K
8K
10K
20
10
8
37.5
18.7
15
39.8
19.9
15.9
0..08
0.04
0.032
100K
0.8
1.5
1.59
0.0032
80K
100K
3