If one of the main parameter and secondary parameter of NG is not qualified, it will be
displayed OUT as shown in Fig.4-7
The comparison process is as shown in Fig.4-8:
Fig.4-7
Note: If you want to save the currently set tolerance value, press the
saving, the data set will not be lost after shutdown.

4.2.11 List Scanning Function

Turn on the instrument to enter the interface of measurement display, move the cursor to the
list switch through the left and right direction keys, and enter Interface of List Scanning by means
of switching the up and down switch Keys as shown in Fig.4-9 and 4-10. The List Scanning
Function can carry out circulated scanning over seven groups of frequencies and compare them
with the nominal values to get the comparison results. The scanning mode is divided into stepping
scanning and sequential scanning. Set the frequency and upper and lower limits, and then open the
scan display, you can get the measured value and comparison results. Comparison result "H":
greater than upper limit, "L": less than lower limit, "I": within the range between them.
Fig.4-9

4.2.12 DCR MODE

Press
to enter DCR DC resistance measurement mode. The interface is as shown in
Fig.4-11.
Fig.4-8
Fig.4-10