Description Of Equipment; Electrical Specification; Supply Requirements; Output Specifications - T&R KV6-80 Operating And Maintenance Manual

High voltage ac test set
Table of Contents

Advertisement

1

DESCRIPTION OF EQUIPMENT

1.1

Electrical Specification

1.1.1

Supply Requirements

KV5-100 mk2 115/230V ±10% 600VA max
KV3-250 mk2 115/230V ±10% 900VA max
KV4-300 mk2 230V ±10% 1400VA max
KV5-200 mk2 230V ±10% 1100VA max
KV6-80
1.1.2

Output Specifications

The output voltage, current and duty cycle are as follows:
Unit
KV5-100 mk2
KV3-250 mk2
KV4-300 mk2
KV5-200 mk2
KV6-80
One side of the high voltage transformer output is connected to the earth system of the
equipment.
1.1.3

Zero Voltage Interlock

The equipment is fitted with a zero volt interlock system on the output control. This interlock
prevents the output being energised unless the output control is in the minimum output
position.
1.1.4

External Interlock

The equipment is fitted with a external interlock system.
panel must have a short circuit across it terminals to enable the output to be energised.
Note: the open circuit voltage across the interlock terminals is 240Vac and is connected
directly to the mains supply.
115/230V ±10% 600VA max
Output voltage
0-5kV
0-3kV
0-4kV
0-5kV
0-6kV
Output current
5 min on/
1 min on/
15 min off
15 min off
100mA
125mA
200mA
100mA
80mA
The interlock socket on the back
7
40 seconds on/
15 min off
-
250mA
300mA
300mA
200mA
-
-
-
-
-