TC6XX-TM6XX Series
 Cyclic ramp configuration?
The figure below illustrates the type of single ramp that can be generated and their parameters:
Programming parameters in Cyclic ramp mode
NH
NB
T0
Tb
Tm
Td
To : time pressing generation start key
Td : Time frame
Tb : Low duration
Tm : Rise (rise time)
Th : High duration
Tf : Drop (drop time)
NB : Low Level
NH : High Level
The LOW level and HIGH level levels are expressed:
As a percentage of the gauge if the scaling mode is ON.
In Volt or in temperature units if the scaling mode is OFF and according to the type of value emitted (voltage or
temperature emission).
The Low Duration, Rise, High Duration, Drop and Time frame durations are expressed in seconds. The max duration is
limited to 1000s.
The Repetitions field indicates the number of ramps that need to be generated. The number of repetitions is limited to
1000.
Use the F2 key to access the CONFIGURATION/RAMP menu. Attention: the appropriate function mode must be
programmed (Single Ramp mode) to access the CONFIGURATION/RAMP/CYCLIC RAMP menu.
In the CONFIGURATION/RAMP/CYCLIC RAMP menu, use the F2 key to go to the next field.
Use the navigation keys to enter the value:
As a percentage of the gauge if the scaling mode is ON.
and  to increase or decrease the value
And  to choose the hundreds/tens/units/decimal place/hundreds/thousands.
Press VAL to save the parameters.
To quit the menu without saving, press CLEAR.