Wahl TM602 User Manual page 31

Pocket rtd and thermocouple temperature meters and calibrators
Table of Contents

Advertisement

 Single ramp signal configuration?
The figure below shows the type of single ramp that can be generated along with its parameters:
Programming parameters in Single Ramp mode
NH
NB
T0
Tdt
To : time pressing the generation start key
Tdt : Total Duration
Tdt : Total Time
NB : Low Level
NH : High Level
The LOW level and HIGH level levels are expressed:
As a percentage of the range if the scaling mode is ON.
In Volt or in temperature units if the scaling mode is OFF and according to the type of value emitted (voltage or
temperature emission).
The Total Duration corresponds to the incrementation time it takes to go from Low Level to High Level (and vice versa
with decrementation). It is given in seconds and is limited to a maximum of 1000s.
The Time frame corresponds to the amount of time you can have between pressing the emission start key and the actual
starting of generation. It is expressed in seconds and the max time is limited to 1000s.
Use the F2 key to access the CONFIGURATION/RAMP menu. Attention: the appropriate function mode must be
programmed (INCREASES mode) to access the CONFIGURATION/RAMP/SINGLE menu.
In the CONFIGURATION/RAMP/SINGLE menu, use the F2 key to go to the next field.
Use the navigation keys to enter the value:
As a percentage of the gauge if the scaling mode is ON.
and  to increase or decrease the value
And  to choose the hundreds/tens/units/decimal place/hundreds/thousands.
Press VAL to save the parameters.
To quit the menu without saving, press CLEAR.
Td
T0
Tdt
Storage of acquisitions in progress
Td
T0
Tdt
D
31