Wahl TM602 User Manual page 14

Pocket rtd and thermocouple temperature meters and calibrators
Table of Contents

Advertisement

TC6XX-TM6XX Series
B.1.7.3
Electrical characteristics not to be exceeded for "voltage" gauges (TC621, TM630 and TM602)
Function
Gauge
U measurement
100mV
U emission (TC6621)
80mV
B.1.7.4
Resistances / temperature measurement (TC622, TM630 and TM612)
Gauge
Resolution (display)
Scope of range:
Scaling
B.1.7.5
Resistance / temperature simulation (TC622)
Gauge
Resolution (display)
Scope of range
Scaling
Measurement current
(Continuous)
Measurement current
(Pulsed)
B.1.7.6
Electrical characteristics not to be exceeded for "resistance" gauges (TC6622, TM630 and TM612)
Function
Range
Ohm measurement
400 Ω / 3600 Ω
Ohm Simulation
400 Ω / 3500 Ω
(TC622 only)
14
Max Vin
60 V
400 Ohm
(for PT100)
10 mΩ or 0.01°
0 Ω to 400 Ω
-220°C to 850°C
-364°F to 1562 °F
yes
400 Ohm
(for PT100)
1 mΩ or 0.01°
0 Ω to 400 Ω
-220°C to 850°C
-364°F to 1562 °F
yes
0.1 mA to 1 mA and 1 mA to 4 mA
0.2 mA to 1 mA and 1 mA to 4 mA
Z load
1000  min
3600 Ohm
(for PT1000)
100 mΩ or 0.01°
0 Ω to 3600 Ω
-220°C to 760°C
-364°F to 1400°F
yes
3500 Ohm
(for PT1000)
1 mΩ or 0.01°
0 Ω to 3500 Ω
-220°C to 715°C
-364°F to 1319°F
yes
0.1mA to 1mA
0.2mA to 1mA
Vin max
Measuring current
60 V
5 mA
USING THE INSTRUMENT

Advertisement

Table of Contents
loading

This manual is also suitable for:

Tm612(a)Tm630(a)Tc621Tc622(a)Tm612Tm630 ... Show all